FDA Enforcement Class II Ongoing

Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluids from a container into the patients vascular system through a vascular access device.

Recall: Z-0793-2020 · Reported January 15, 2020

Enforcement

Recall Number
Z-0793-2020
Event ID
84511
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 15, 2020
Initiation Date
December 12, 2019
Classification Date
January 9, 2020
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluids from a container into the patients vascular system through a vascular access device.

Reason

Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.

Code Info

Product Codes: 2C8519, UDI 00085412048994; 2C8537, UDI 00085412048970; and 2C8541, UDI 00085412071107 All Lots

Distribution

Worldwide distribution - US Nationwide distribution and countries of Bermuda, Bahamas, Trinidad and Tobago, Grand Cayman, St. Thomas, Guyana and Barbados, Mexico, Canada, and New Zealand.

Quantity

271,014,984 units