FDA Enforcement
Class III
Terminated
OPTIGEN Food 30; Catalog number: 85035. For allergen testing. Manufactured by Hitachi Chemical Diagnostics, Mountain View, CA. Distributed by Theridiag, Cedex, France
Recall: Z-0793-2013
·
Reported February 13, 2013
Enforcement
- Recall Number
- Z-0793-2013
- Event ID
- 64132
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hitachi Chemical Diagnostics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- February 13, 2013
- Initiation Date
- December 21, 2012
- Classification Date
- February 7, 2013
- Termination Date
- May 16, 2013
- Address
- 630 Clyde Ct, N/A, Mountain View, CA, 94043-2239, United States
Description
OPTIGEN Food 30; Catalog number: 85035. For allergen testing. Manufactured by Hitachi Chemical Diagnostics, Mountain View, CA. Distributed by Theridiag, Cedex, France
Reason
There is no evidence that a visual inspection was performed on one tray of coverslips. Possibility of a low fill volume which could lead to a false negative results.
Code Info
Catalog number: 85035. Lot number 42K283626, expiration May 31, 2013.
Distribution
Worldwide Distribution - France only
Quantity
20 kits