FDA Enforcement Class II Terminated

Mizuho OSI Trios, Table Base; Voltage: 100-240V ~ 50/60 Hz Amps: 5,0-5.0A; Model 7803 and 7803AP Orthopaedic: Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.

Recall: Z-0792-2016 · Reported February 17, 2016

Enforcement

Recall Number
Z-0792-2016
Event ID
72937
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mizuho OSI
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
February 17, 2016
Initiation Date
December 14, 2015
Classification Date
February 9, 2016
Termination Date
February 10, 2016
Address
30031 Ahern Ave, N/A, Union City, CA, 94587-1234, United States

Description

Mizuho OSI Trios, Table Base; Voltage: 100-240V ~ 50/60 Hz Amps: 5,0-5.0A; Model 7803 and 7803AP Orthopaedic: Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.

Reason

Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.

Code Info

Model numbers 7803, 7803AP. Serial numbers 228-261.

Distribution

US, South Africa, Japan and United Arab Emirates.

Quantity

33