FDA Enforcement
Class II
Terminated
3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair and head hair from patients in preparation for any medical procedure requiring hair removal.
Recall: Z-0791-2019
·
Reported February 6, 2019
Enforcement
- Recall Number
- Z-0791-2019
- Event ID
- 81933
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- 3M Company - Health Care Business
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 6, 2019
- Initiation Date
- January 14, 2019
- Classification Date
- January 30, 2019
- Termination Date
- February 11, 2021
- Address
- 3m Center 2510 Conway Ave , B# 275-5-W-6, N/A, Saint Paul, MN, 55144-0001, United States
Description
3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair and head hair from patients in preparation for any medical procedure requiring hair removal.
Reason
3M has recently received reports involving the 3MTM Surgical Clipper by Remington¿, Model 9602, overheating in mobile/field EMS situations while charging from a modified sine wave power inverter.
Code Info
all lots
Distribution
US Distribution and countries Brazil and Canada.
Quantity
19,860 units