FDA Enforcement Class II Terminated

Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur in order to create space for the femoral stem during a hip replacement surgery.

Recall: Z-0791-2013 · Reported February 13, 2013

Enforcement

Recall Number
Z-0791-2013
Event ID
64160
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medacta Usa Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 13, 2013
Initiation Date
December 17, 2012
Classification Date
February 6, 2013
Termination Date
February 7, 2013
Address
4725 Calle Quetzal Ste B, N/A, Camarillo, CA, 93012-8429, United States

Description

Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur in order to create space for the femoral stem during a hip replacement surgery.

Reason

The tip of the Quadra Trial Broach size 0, broke during a surgery.

Code Info

Lot/Unit Codes: 075023, 075064, 084000, 084001, 084326, 084743, 084815, 085202, 085455

Distribution

US Nationwide Distribution including the states of Arizona, California, Idaho, Illinois, North Carolina, Ohio, Utah, Washington and US Virgin Islands

Quantity

26 pieces U.S.A. (226 pieces Worldwide)