FDA Enforcement Class II Terminated

Vis-U-All High Temp 4.5"x11" Heat Seal Pouch 200 pouches per box; 5 boxes per case Model Number: 883411

Recall: Z-0790-2021 · Reported January 13, 2021

Enforcement

Recall Number
Z-0790-2021
Event ID
86942
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Steris Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 13, 2021
Initiation Date
December 2, 2020
Classification Date
January 7, 2021
Termination Date
June 21, 2022
Address
5960 Heisley Rd, Mentor, OH, 44060-1834, United States

Description

Vis-U-All High Temp 4.5"x11" Heat Seal Pouch 200 pouches per box; 5 boxes per case Model Number: 883411

Reason

Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch

Code Info

Lot Number: 200421A, 200521A

Distribution

Nationwide Foreign:

Quantity

53 cases