FDA Enforcement Class II Terminated

Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302.

Recall: Z-0788-2020 · Reported January 15, 2020

Enforcement

Recall Number
Z-0788-2020
Event ID
84227
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Avanos Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
January 15, 2020
Initiation Date
November 5, 2019
Classification Date
January 8, 2020
Termination Date
August 16, 2023
Address
5405 Windward Pkwy, N/A, Alpharetta, GA, 30004-3894, United States

Description

Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302.

Reason

Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.

Code Info

REF 195 Lot Code / Expiration Date: M18115T401/2023-APR, M18121T402/2023-MAY, M18128T404/2023-MAY, M18143T401/2023-MAY, M18149T403/2023-MAY, M18156T402/2023-JUN, M18170T402/2023-JUN, M18177T402/2023-JUN, M18192T402/2023-JUL, M18206T401/2023-JUL, M18220T402/2023-AUG, M18247T402/2023-SEP, M18254T405/2023-SEP, M18282T404/2023-OCT, M18311T401/2023-NOV, M18318T402/2023-NOV, M18345T403/2023-DEC, M1834T403/2023-DEC, M18358T403/2023-DEC, M19008T401/2024-JAN, M19058T402/2024-FEB, M19078T403/2024-MAR, M19086T401/2024-MAR, M19098T403/2024-APR, M19134T404/2024-MAY, REF 201 Lot Code / Expiration Date: M18044T405/2023-FEB, M18311T402/2023-NOV, M18358T404/2023-DEC, M19037T402/2024-FEB, M19065T402/2024-MAR, M7172T402/2022-JUN. REF 202 Lot Code / Expiration Date: M18143T403/2023-MAY, M18192T401/2023-JUL, M18254T404/2023-SEP, M18274T401/2023-OCT, M18296T405/2023-OCT, M18318T401/2023-NOV, M19072T401/2024-MAR, M19142T401/2024-MAY, M7346T403/2022-DEC. REF 195-5 Lot Code / Expiration Date: 20000095/2024 - JUL, M18274T402/2023-OCT, M18290T402/2023-OCT, M18296T406/2023-OCT, M18311T403/2023-NOV, M18324T401/2023-NOV, M18324T402/2023-NOV, M18332T401/2023-NOV, M18339T401/2023-DEC, M18345T404/2023-DEC, M18352T402/2023-DEC, M19024T402/2024-JAN, M19030T401/2024-JAN, M19051T401/2024-FEB, M19059T401/2024-FEB, M19065T401/2024-MAR, M19072T402/2024-MAR, M19079T401/2024-MAR, M19085T402/2024-MAR, M19098T404/2024-APR, M19099T401/2024-APR, M19105T403/2024-APR, M19120T403/2024-APR, M19128T401/2024-MAY, M19128T403/2024-MAY, M19135T401/2024-MAY, M19142T402/2024-MAY, M19149T405/2024-MAY, M19162T403/2024-JUN, M19177T402/2024-MAY, M19184T401/2024-JUL.

Distribution

Nationwide. Internationally distributed to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Honduras, India, Ireland, Israel, Italy, Korea, Republic of, Lebanon, Mexico, Nepal, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.

Quantity

114,300 units (37,880 US, 76,420 OUS)