FDA Enforcement Class II Terminated

SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction.

Recall: Z-0787-2016 · Reported February 17, 2016

Enforcement

Recall Number
Z-0787-2016
Event ID
73076
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
DePuy Orthopaedics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 17, 2016
Initiation Date
January 8, 2016
Classification Date
February 9, 2016
Termination Date
March 24, 2017
Address
700 Orthopaedic Dr, N/A, Warsaw, IN, 46582-3994, United States

Description

SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction.

Reason

Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch dislodges from the inserter and needs to be removed from the joint space. Insert Dislocation or Spin Out, or Poor Joint Mechanics may result, if not noted during surgery. May require revision surgery.

Code Info

Cat. No. GTIN Lot Description 950502016 10603295227366 ABB92402 SIGMA HP MBT Cemented Punch size 1-1.5 950502017 10603295227373 ABB84491 SIGMA HP MBT Cemented Punch size SZ 2-3 950502018 10603295227380 ABB92358 SIGMA HP MBT Cemented Punch size SZ 4-7 950502020 10603295227403 ABC33125 SIGMA HP MBT Noncemented Punch size 2-3 950502020 10603295227403 ABB81733 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502020 10603295227403 ABB81732 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502021 10603295227410 ABB39952 SIGMA HP MBT Noncemented Punch size SZ 4-7 950502021 10603295227410 ABC33123 SIGMA HP MBT Noncemented Punch size SZ 4-7

Distribution

Worldwide Distribution-US Nationwide, Armenia, Australia, Austria, Belgium, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, P R China, Portugal, Poland, Russia, Singapore, Slovenia, South Africa, Spain, Sweden, South Korea, Switzerland, Thailand, Turkey, UAE, UK, and Brazil.

Quantity

351