FDA Enforcement Class II Terminated

PRIMUS, ONCOR, ARTISTE Linac systems with component 550 TxT Treatment Table. TT-A, TT-D, TT-M or TT-S Table tops. Manufactured by: SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse 19-21 Kemnath, GERMANY 95478 The intended use of the SIEMENS branded, ARTISTE", ONCOR" and PRIMUS" family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Recall: Z-0787-2014 · Reported January 29, 2014

Enforcement

Recall Number
Z-0787-2014
Event ID
67126
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2014
Initiation Date
December 5, 2013
Classification Date
January 17, 2014
Termination Date
January 30, 2015
Address
757 Arnold Dr Ste A, Martinez, CA, 94553-3615, United States

Description

PRIMUS, ONCOR, ARTISTE Linac systems with component 550 TxT Treatment Table. TT-A, TT-D, TT-M or TT-S Table tops. Manufactured by: SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse 19-21 Kemnath, GERMANY 95478 The intended use of the SIEMENS branded, ARTISTE", ONCOR" and PRIMUS" family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Reason

Update to correct the possibility for patients or operators to have their fingers trapped or pinched between the table top and the bottom frame of the 550 TxT Treatment Table. Combined with this update instruction is a second complaint issue related to an incident in which a cable inside of the TxT" Treatment Table has been damaged due to contacting a moving part. The resulting short-circuit cau

Code Info

550 TxT table - part number 07346534.

Distribution

Worldwide Distribution: US (nationwide) and internationally to: Canada, Italy, P.R. China, Argentina, Pakistan, South Africa, Egypt, Iran, Vietnam, Indonesia, Brazil, Hungary, Bulgaria, India, Spain, Switzerland, Trinidad, Tobago, Italy, Algeria, Croatia, Turkey, Japan, Germany, Portugal, Poland, France, Sri Lanka, Russian Federation, France, Lebanon, Slovakia, Ukraine, Pakistan, and Korean Republic.

Quantity

305 active devices.