Mx8000 IDT Whole Body Computed Tomography X-ray system Distributed by Philips Medical Systems, Highland Heights, OH 44143 The Brilliance Scanner are computed Tomography X-ray systems with multi-slice capability of up to 16 slices,which are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Enforcement
- Recall Number
- Z-0786-2013
- Event ID
- 63392
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems (Cleveland) Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 6, 2013
- Initiation Date
- September 18, 2012
- Classification Date
- February 26, 2013
- Termination Date
- September 2, 2015
- Address
- 595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States
Description
Mx8000 IDT Whole Body Computed Tomography X-ray system Distributed by Philips Medical Systems, Highland Heights, OH 44143 The Brilliance Scanner are computed Tomography X-ray systems with multi-slice capability of up to 16 slices,which are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
The low contrast pins in the Image Quality (IQ) phantom deteriorates over time. This prevents the customers from doing the Daily Image Quality Checks as defined in the current Instructions for Use (IfU).
1. Brilliance CT 16 slice (water): System Code #, 728245, Serial #s:2506, 2518, 2555, 2558, 2559, 2561, 2561, 2562, 2564, 2567, 2573, 2576, 2578, 2582, 2583, 2584, 2585, 2586, 2587, 2589, 2590, 2592, 2594, 2596, 2597, 3014, 3051, 3090, 3091, 3105, 3151, 3217, 3235, 3272, 3287, 3303, 3312, 3335, 3338, 3339, 3340, 3341, 3342, 3343, 3344, 3345, 3346, 3347, 3348, 3349, 3350, 3351, 3352, 3353, 3354, 3355, 3356, 3357, 3358, 3359, 3359, 3360, 3361, 3362, 3363, 3364, 3365, 3366, 3367, 3368, 3369, 3370, 3371, 3372, 3373, 3374, 3375, 3376, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3384, 3385, 3386, 3387, 3388, 3389, 3390, 3391, 3392, 3393, 3394, 3395, 3396, 3397, 3398, 3399, 3400, 3401, 3402, 3403, 3404, 3405, 3406, 3407, 3408, 3409, 3410, 3411, 3412, 3413, 3414, 3415, 3416, 3417, 3418, 3419, 3420, 3421, 3422, 3423, 3424, 3425, 3426, 3427, 3428, 3429, 3430, 3431, 3432, 3433, 3434, 3435, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3442, 3443, 3444, 3445, 3446, 3447, 3448, 3449, 3450, 3451, 3452, 3453, 3454, 3455, 3456, 3457, 3458, 3459, 3460, 3461, 3462, 3463, 3464, 3465, 3466, 3467, 3468, 3469, 3470, 3471, 3472, 3473, 3474, 3475, 3476, 3477, 3478, 3479, 3480, 3481, 3482, 3483, 3484, 3485, 3486, 3487, 3488, 3489, 3490, 3491, 3492, 3493, 3494, 3495, 3496, 3497, 3498, 3499, 3500, 3501, 3502, 3503, 3504, 3505, 3506, 3507, 3508, 3509, 3510, 3511, 3512, 3513, 3514, 3515, 3516, 3517, 5010, 5077, 5119, 6095, 9701, 9708, 9715, 9716, 9718, 9719, 9724, 9728, 9729, 9732, 9741, 9743, 9745, 9746, 9749, 9750, 9754, 9760, 9764, 9771, 9776, 9778, 9780, 9785, 9787, 9788, 9789, 9790, 9794, 9797, 9800, 9801, 9806, 9808, 9810, 9999 & 111111. 2. Brilliance CT 10 slice (water): System Code #, 728250, Serial #s: 2554, 2555, 2556, 2557, 2558, 2559, 2560, 2561, 2562, 2563, 2564, 2565, 2566, 2567, 2568, 2569, 2570, 2571, 2572, 2573, 2574, 2575, 2576, 2577, 2578, 2579, 2580, 2581, 2582, 2583, 2585, 2586, 2587, 2588, 2589, 2590, 2591, 2592, 2593, 2594, 2595, 2596, 2597, 2598, 9704, 9708, 9738 & 9741. 3. Brilliance CT 6 slice (water): System Code #, 728255, Serial #s: 2565, 9701, 9702, 9703, 9704, 9705, 9706, 9707, 9708, 9709, 9710, 9711, 9712, 9713, 9714, 9715, 9716, 9717, 9718, 9719, 9720, 9721, 9722, 9723, 9724, 9725, 9726, 9727, 9728, 9729, 9730, 9731, 9732, 9733, 9734, 9735, 9736, 9737, 9738, 9739, 9740, 9741, 9742, 9743, 9744, 9745, 9746, 9747, 9748, 9749, 9750, 9751, 9752, 9753, 9754, 9755, 9756, 9757, 9758, 9759, 9760, 9761, 9762, 9763, 9764, 9765, 9766, 9767, 9768, 9769, 9770, 9771, 9772, 9773, 9774, 9775, 9776, 9777, 9778, 9779, 9780, 9781, 9782, 9783, 9784, 9785, 9786, 9787, 9788, 9789, 9790, 9791, 9792, 9793, 9794, 9795, 9796, 9797, 9798, 9799, 9800, 9801, 9802, 9803, 9804, 9805, 9806, 9807, 9808, 9810 & 111111. 4. Mx8000 IDT 16 slice System Code #, 728120, Serial #, Serial #s: 1225, 2059, 2345, 2501, 2502, 2506, 2510, 2511, 2513, 2516, 2517, 2518, 2520, 2525, 2528, 2531, 2533, 2535, 2537, 2539, 2546, 2550, 3002, 3006, 3007, 3008, 3012, 3013, 3014, 3017, 3019, 3020, 3021, 3022, 3023, 3024, 3025, 3026, 3027, 3028, 3029, 3030, 3031, 3032, 3034, 3035, 3036, 3037, 3038, 3039, 3040, 3041, 3042, 3043, 3044, 3045, 3046, 3047, 3048, 3049, 3050, 3051, 3052, 3053, 3054, 3055, 3056, 3057, 3058, 3059, 3060, 3061, 3062, 3063, 3064, 3065, 3066, 3067, 3068, 3069, 3070, 3071, 3072, 3073, 3074, 3075, 3076, 3077, 3078, 3079, 3082, 3083, 3084, 3085, 3086, 3087, 3088, 3089, 3090, 3091, 3092, 3093, 3094, 3095, 3096, 3097, 3098, 3099, 3100, 3101, 3102, 3103, 3104, 3105, 3106, 3107, 3108, 3109, 3110, 3111, 3112, 3113, 3114, 3115, 3116, 3117, 3118, 3119, 3120, 3121, 3122, 3123, 3124, 3125, 3126, 3127, 3128, 3129, 3130, 3131, 3132, 3133, 3134, 3135, 3136, 3137, 3138, 3139, 3140, 3141, 3143, 3144, 3145, 3146, 3147, 3148, 3149, 3150, 3151, 3152, 3153, 3154, 3155, 3156, 3157, 3158, 3159, 3160, 3161, 3162, 3163, 3164, 3165, 3167, 3168, 3169, 3170, 3171, 3172, 3173, 3174, 3175, 3176, 3177, 3178, 3179, 3180, 3181, 3182, 3183, 3184, 3185, 3186, 3187, 3188, 3189, 3190, 3191, 3192, 3193, 3194, 3195, 3196, 3197, 3198, 3199, 3200, 3201, 3202, 3203, 3204, 3205, 3206, 3207, 3208, 3209, 3210, 3211, 3212, 3213, 3214, 3215, 3216, 3217, 3218, 3219, 3220, 3221, 3222, 3223, 3224, 3225, 3226, 3227, 3228, 3229, 3230, 3231, 3232, 3233, 3234, 3235, 3236, 3237, 3238, 3239, 3240, 3241, 3242, 3243, 3244, 3245, 3246, 3247, 3248, 3249, 3250, 3251, 3252, 3253, 3254, 3255, 3256, 3257, 3258, 3259, 3260, 3261, 3262, 3263, 3264, 3265, 3266, 3267, 3268, 3269, 3270, 3271, 3272, 3273, 3274, 3275, 3276, 3277, 3278, 3279, 3280, 3281, 3282, 3283, 3284, 3285, 3286, 3287, 3288, 3289, 3290, 3291, 3292, 3293, 3294, 3295, 3296, 3297, 3298, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3307, 3308, 3309, 3310, 3311, 3312, 3313, 3314, 3315, 3316, 3317, 3318, 3319, 3320, 3321, 3322, 3323, 3324, 3325, 3326, 3327, 3328, 3329, 3330, 3333, 3336, 3367, 3420, 3480, 4111, 4112, 4114, 4115, 4116, 4117, 4444, 4445, 4446, 4447, 4567, 4799, 5005, 5006, 5007, 5008, 5009, 5011, 5012, 5013, 5014, 5555, 5678, 8888, 9760, 9789, 9900, 9991, 9995, 9997, 9999, 11111, 80159 & 111111. 5. Mx8000 IDT 10 slice: System Code #, 728125, Serial #s: 1225, 1234, 2501, 2502, 2503, 2504, 2505, 2506, 2507, 2508, 2509, 2510, 2511, 2512, 2513, 2514, 2515, 2516, 2517, 2518, 2519, 2520, 2521, 2522, 2523, 2524, 2525, 2526, 2527, 2528, 2529, 2530, 2531, 2532, 2533, 2534, 2535, 2536, 2537, 2538, 2539, 2540, 2541, 2542, 2543, 2544, 2546, 2547, 2548, 2549, 2550, 2551, 2552, 3018, 9999 & 111111. 6. Mx8000 Quad: System Code #, 728200, Serial #s: 2008, 2009, 2011, 2014, 2017, 2022, 2023, 2025, 2027, 2029, 2030, 2031, 2032, 2034, 2035, 2037, 2038, 2041, 2042, 2044, 2045, 2046, 2047, 2048, 2050, 2051, 2055, 2056, 2058, 2059, 2061, 2062, 2063, 2064, 2065, 2068, 2073, 2074, 2075, 2080, 2081, 2084, 2085, 2086, 2089, 2092, 2093, 2095, 2096, 2099, 2100, 2101, 2102, 2103, 2104, 2105, 2109, 2116, 2123, 2125, 2126, 2127, 2128, 2129, 2131, 2132, 2134, 2135, 2136, 2137, 2138, 2139, 2140, 2141, 2142, 2150, 2151, 2152, 2155, 2156, 2158, 2159, 2161, 8118, 9124, 204935 & 2020-1. 7. Mx8000 Dual: System Code #, 728201, Serial #s: 4288, 9000, 9005, 9006, 9008, 9010, 9011, 9012, 9015, 9018, 9020, 9021, 9025, 9027, 9035, 9036, 9037, 9038, 9039, 9042, 9049, 9050, 9055, 9056, 9058, 9059, 9061, 9062, 9063, 9068, 9072, 9076, 9077, 9080, 9084, 9085, 9089, 9090, 9091, 9093, 9094, 9096, 9098, 9103, 9105, 9106, 9108, 9109, 9110, 9120, 9121, 9125, 9131, 9133, 9135, 9136, 9137, 9138, 9139, 9143, 9144, 9146, 9148, 9151, 9155, 9160, 9163, 9169, 9173, 9174, 9175, 9179, 9182, 9184, 9187, 9188, 9194, 9195, 9196, 9197, 9198, 9199, 9200, 9201, 9202, 9205, 9208, 9210, 9211, 9212, 9219, 9220, 9222, 9228, 9231, 9232, 9239, 9240, 9241, 9242, 9249, 9250, 9254, 9257, 9258, 9259, 9260, 9261, 9262, 9266, 9267, 9269, 9270, 9271, 9272, 9273, 9274, 9278, 9279, 9280, 9281, 9284, 9285, 9288, 9290, 9292, 9294, 9295, 9296, 9297, 9299, 9300, 9301, 9307, 9308, 9317, 9320, 9324, 9327, 9328, 9331, 9332, 9334, 9336, 9339, 9351, 9352, 9354, 9357, 9361, 9362, 9363, 9364, 9365, 9367, 9368, 9369, 9370, 9372, 9376, 9377, 9378, 9380, 9384, 9387, 9396, 9397, 9405, 9407, 9411, 9412, 9417, 9423, 9426, 9427, 9428, 9429, 9431, 9433, 9445, 9446, 9447, 9465, 9466, 9467, 9469, 9471, 9472, 9473, 9474, 9476, 9477, 9482, 9483, 9484, 9487, 9488, 9489, 9490, 9493, 9496, 9497, 9500, 9509, 9510, 9515, 9521, 9522, 9529, 9530, 9532, 9537, 9538, 9539, 9540, 9541, 9542, 9543, 9544, 9549, 9551, 9552, 9553, 9555, 9561, 9563, 9574, 9576, 9577, 9584, 9585, 9588, 9592, 9593, 9598, 9599, 9600, 9601, 9603, 9606, 225637, 226925, 235225, 238067, 999030, 9122A, 9347A, 9361-1 & 9586-1. 8. Mx8000 Dual exp: System Code #, 728130, Serial #, Serial #s: 2110, 3291, 6005, 8008, 8010, 8014, 8016, 8018, 8021, 8023, 8025, 8034, 8035, 8039, 8040, 8041, 8043, 8046, 8048, 8053, 8054, 8055, 8056, 8057, 8058, 8061, 8064, 8065, 8066, 8067, 8068, 8070, 8073, 8074, 8075, 8076, 8077, 8078, 8079, 8080, 8081, 8082, 8083, 8084, 8085, 8086, 8091, 8092, 8095, 8096, 8097, 8098, 8102, 8103, 8104, 8105, 8106, 8107, 8108, 8111, 8112, 8113, 8114, 8115, 8116, 8117, 8118, 8119, 8120, 8122, 8123, 8124, 8125, 8126, 8127, 8128, 8129, 8130, 8131, 8132, 8134, 8135, 8136, 8137, 8138, 8139, 8140, 8142, 8143, 8144, 8145, 8146, 8147, 8149, 8150, 8151, 8152, 8153, 8154, 8155, 8156, 8157, 8158, 8159, 8160, 8162, 8163, 8164, 8165, 8166, 8167, 8168, 8169, 8171, 8172, 8173, 8174, 8175, 8176, 8177, 8178, 8180, 8181, 8183, 8184, 8185, 8186, 8187, 8188, 8190, 8191, 8192, 8193, 8194, 8195, 8196, 8197, 8198, 8199, 8200, 8201, 8202, 8203, 8204, 8205, 8206, 8207, 8208, 8210, 8211, 8212, 8213, 8215, 8216, 8217, 8218, 8219, 8220, 8221, 8222, 8223, 8224, 8225, 8226, 8227, 8228, 8229, 8230, 8232, 8233, 8234, 8235, 8236, 8237, 8238, 8239, 8240, 8242, 8243, 8244, 8245, 8246, 8247, 8248, 8249, 8250, 8253, 8254, 8255, 8256, 8257, 8258, 8259, 8260, 8261, 8262, 8263, 8264, 8265, 8266, 8267, 8268, 8269, 8270, 8271, 8272, 8273, 8274, 8275, 8276, 8277, 8278, 8279, 8280, 8281, 8282, 8283, 8284, 8285, 8286, 8287, 8288, 8289, 8290, 8291, 8292, 8293, 8294, 8295, 8296, 8297, 8298, 8299, 8300, 8301, 8302, 8303, 8304, 8305, 8306, 8307, 8308, 8309, 8310, 8311, 8312, 8313, 8314, 8315, 8316, 8317, 8318, 8319, 8320, 8321, 8322, 8323, 8324, 8325, 8326, 8327, 8328, 8329, 8330, 8331, 8332, 8333, 8334, 8335, 8336, 8337, 8338, 8339, 8340, 8341, 8342, 8343, 8344, 8345, 8346, 8347, 8349, 8350, 8351, 8352, 8353, 8354, 8355, 8356, 8357, 8358, 8359, 8360, 8361, 8362, 8363, 8364, 8365, 8366, 8367, 8368, 8369, 8370, 8371, 8372, 8373, 8374, 8375, 8376, 8377, 8378, 8379, 8380, 8381, 8382, 8383, 8384, 8385, 8386, 8387, 8388, 8389, 8390, 8391, 8392, 8393, 8394, 8395, 8396, 8398, 8400, 8401, 8402, 8403, 8404, 8405, 8406, 8407, 8409, 8410, 8411, 8412, 8413, 8442, 9528, 80002, 80003, 80004, 80005, 80007, 80009, 80010, 80011, 80012, 80013, 80014, 80015, 80016, 80017, 80018, 80019, 80021 & 80022.
Worldwide Distribution - USA ( nationwide) including the states of AK, AL, AR, AZ, CA, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, VA, WA, WV and WY, and the countries of Albania, Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Brazil, Canada, China, Colombia, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Ghana, Gibraltar, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Venezuela, Viet Nam & Yemen.
1,474 Units