FDA Enforcement Class II Ongoing

iQ200 Series Urine Microscopy Analyzers, Part Numbers: a) 700-3320, b) 700-3370, c) 700-3345, d) 700-3347, e) C10683, f) 700-3375, g) 700-3325, h) C10684

Recall: Z-0785-2024 · Reported January 31, 2024

Enforcement

Recall Number
Z-0785-2024
Event ID
93655
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 31, 2024
Initiation Date
December 19, 2023
Classification Date
January 25, 2024
Address
11800 Sw 147th Ave, N/A, Miami, FL, 33196-2500, United States

Description

iQ200 Series Urine Microscopy Analyzers, Part Numbers: a) 700-3320, b) 700-3370, c) 700-3345, d) 700-3347, e) C10683, f) 700-3375, g) 700-3325, h) C10684

Reason

Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

Code Info

a) 700-3320, UDI/DI 10837461001188; b) 700-3370, UDI/DI 10837461001324; c) 700-3345, UDI/DI 10837461001300; d) 700-3347, UDI/DI 10837461001317; e) C10683, UDI/DI 15099590697129 f) 700-3375, UDI/DI 10837461001355; g) 700-3325, UDI/DI 10837461001256; h) C10684, UDI/DI 15099590695484; All serial numbers

Distribution

US and Australia, Brazil, Canada, China, France, Germany, Ghana, India, Israel, Italy, Lebanon, Lithuania, Mexico, Panama, Qatar, Romania, Saudi Arabia, Spain, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland,

Quantity

112 units