FDA Enforcement Class II Terminated

ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

Recall: Z-0785-2014 · Reported January 29, 2014

Enforcement

Recall Number
Z-0785-2014
Event ID
67130
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2014
Initiation Date
December 4, 2013
Classification Date
January 17, 2014
Termination Date
November 6, 2015
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

Reason

The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.

Code Info

ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005, 9510027, 9510028, 9603042, 9603043, 9604044, 9604047, 9605011, 9607060, 9609065, 9610072, 9704081, 9709090, 9710097, 9711103, 9712105, 9712106, 9801107, 9801109, 9802112, 9803001, 9803116, 9804126, 9804129, 9806136, 9807138, 9807422,9810527, 9901156, 9901160, 9901162, 9908167, 9909172, 9910176, 00709B, 0101003A, 01310B, 01525B, 01920B, 01921B, 01922B, 03041B, 03094B, 03338B, 03368B, 2145-30074A, 7718B, A0103015, A0103213, A0105217, A02100230, A02100231, A02110248, A02110249, A02110250, A02110251, A02120254, A02120256, A03010262, A03010264, A03010265, A03010266, A03010270, A03020274, A03020275, A03030282, A03070283, A03120294, B0003196, B0007200, B0309014, ID3935B, M@10639, M@10645, M@10774, M@10777, M@10877, M@11239, M@11568, M@11640, M@11684, M@11685, M@11729, M@11769, M@11813, M@11947, M@12378-1, M@12462, M@12765, M@13725, M@24470, M@4750028851, M@4750028853, M@5348B, M@5807B, M@5874B, M@5894B, M@5932B, M@5941B, M@5963B, M@6041B, M@6062B, M@6229B, M@6354B, M@6477B, M@6485B, M@6546B, M@6555B, M@6692B, M@6754B, M@6809B, M@6815B, M@6948B, M@7007B, M@7023B, M@7093B, M@7194B, M@7198B, M@7207B, M@7214B, M@7280B, M@7285B, M@7286B, M@7408B, M@7540B, M@7596B, M@7675B, M@7704B, M@8156B, M@8324B, M@8344B, M@8429B, M@8523B, M@8578B, M@8588B, M@8637B, M@8639RB, M@8648B, M@8808B, M@8813B, M@8898B, M@8977B, M@9050B, M@9261B, M@9276B, M@9355RB, M@9475B, M@9544B, M@9647B, M@9656B, M@9733B & M@9796B. ADAC ARGUS: Model #88295; Serial #: 9211014, 9301005, 9304019, 9304024, 9304028, 9401028, 9401031, 9401037, 9505014, 9910033, 93-10678- 1292, 9910028P, M@10225AMT, M@3732B, M@3903B, M@3954B, M@3956B, M@3960B, M@4249B, M@4415B, M@4646B, M@4868B, M@5172B, M@5321B, M@7341B & M@AMT1154. ADAC CIRRUS: Model #88294; 34, 718558, 9206009, 9207003, 9207004, 9207005, 9207007, 9211006, 9300113, 9304013, 9309012, 9404022, 9407028, 9509042, 9604098, 9910030, 93-10735-0393, 9621RQ, M@10110AMT, M@10324AMT, M@10402AMT, M@3599G-08, M@3905Q, M@4029Q, M@4129Q, M@4142Q, M@4668Q, M@4743Q, M@5003Q, M@5668Q, M@5895Q, M@AMT126R, M@AMT1282 & M@MVS3227A.

Distribution

Worldwide Distribution: US (nationwide) including states of: AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great Britain, Italy, Netherlands, Norway, Pakistan, Russia, Singapore & South Korea.

Quantity

255 units