FDA Enforcement Class II Terminated

(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR CARRIER SLIM/RAISED, LR 3101705 TRAV RAIL CARR ESYST WIDE 3102517 TR CARR. NORMAL/LOWER. LR 3102506 TR CARRIER WIDE/RAISED ML 3102519 TR CARR. LOW.LR 30MM/PAIR 3102511 TR CARRIER NORMAL, LR 3102531 TRAVERSE RL CARRIER STD/RAISED 3102512 TR CARRIER WIDE, LR 3102532 TRAVERSE RL CARRIER WD/RAISED 3102513 TR CARRIER SLIM, LR 3102562 TR CARRIER, PARALLEL/WIDE

Recall: Z-0783-2022 · Reported March 23, 2022

Enforcement

Recall Number
Z-0783-2022
Event ID
89562
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hill-Rom, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 23, 2022
Initiation Date
January 14, 2022
Classification Date
March 11, 2022
Termination Date
November 7, 2024
Address
1069 State Road 46 E, N/A, Batesville, IN, 47006-7520, United States

Description

(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR CARRIER SLIM/RAISED, LR 3101705 TRAV RAIL CARR ESYST WIDE 3102517 TR CARR. NORMAL/LOWER. LR 3102506 TR CARRIER WIDE/RAISED ML 3102519 TR CARR. LOW.LR 30MM/PAIR 3102511 TR CARRIER NORMAL, LR 3102531 TRAVERSE RL CARRIER STD/RAISED 3102512 TR CARRIER WIDE, LR 3102532 TRAVERSE RL CARRIER WD/RAISED 3102513 TR CARRIER SLIM, LR 3102562 TR CARRIER, PARALLEL/WIDE

Reason

Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients

Code Info

Distributed between 04JAN2021 and 24MAR2021. P/N Product name 3130001 Multirall 200 3301030 LikoGuard L 3301040 LikoGuard XL 3121001 Likorall 200 3122005 Likorall 242 ES natural 3122005NA Likorall 242 ES natural 3122006 Likorall 242 ES white 3122006NA Likorall 242 ES white 3122009 Likorall 242 S natural 3122009NA Likorall 242 S natural 3122010 Likorall 242 S white 3122010NA Likorall 242 S white 3122007 Likorall 242 ES R2R, natural 3122007NA Likorall 242 ES R2R, natural 3122008 Likorall 242 ES R2R, white 3122008NA Likorall 242 ES R2R, white 3122011 Likorall 242 S R2R, natural 3122011NA Likorall 242 S R2R, natural 3122012 Likorall 242 S R2R, white 3122012NA Likorall 242 S R2R, white 3123001 Likorall 243 ES, natural 3123001NA Likorall 243 ES, natural 3123002 Likorall 243 ES, white 3123002NA Likorall 243 ES, white 3122501 Likorall 250 ES, natural 3122501NA Likorall 250 ES, natural 3122502 Likorall 250 ES, white 3122502NA Likorall 250 ES, white 3124050 Likorall 250 S, IRC, natural

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IT, JP, NL, NO, NZ, RU, SE, SG, SK.

Quantity

Traverse Rail Carrier E-System, wide 31017XX 5 Traverse Rail Carrier 31025XX 1798