Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.
Recall: Z-0783-2013
·
Reported February 13, 2013
Enforcement
- Recall Number
- Z-0783-2013
- Event ID
- 63793
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- The Anspach Effort, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 13, 2013
- Initiation Date
- September 5, 2012
- Classification Date
- February 4, 2013
- Termination Date
- August 13, 2014
- Address
- 4500 Riverside Drive, N/A, Palm Beach Gardens, FL, 33410-4235, United States
Description
Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.
Reason
Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.
Code Info
Lot number: 2000930
Distribution
Distributed in the state of Massachusetts
Quantity
8 devices