FDA Enforcement Class II Terminated

XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.

Recall: Z-0780-2014 · Reported January 29, 2014

Enforcement

Recall Number
Z-0780-2014
Event ID
67078
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Elekta, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2014
Initiation Date
November 25, 2013
Classification Date
January 17, 2014
Termination Date
May 13, 2014
Address
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227, United States

Description

XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.

Reason

When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.

Code Info

Model or Catalog No. MUJ; Software Version XiO Versions 4.80 and higher

Distribution

Worldwide Distribution - US Distribution including the states of IL, IN, LA, NJ, TX, WA and WI., and the countries of Belgium, Germany, Italy and Netherlands.

Quantity

13