XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
Enforcement
- Recall Number
- Z-0780-2014
- Event ID
- 67078
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Elekta, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 29, 2014
- Initiation Date
- November 25, 2013
- Classification Date
- January 17, 2014
- Termination Date
- May 13, 2014
- Address
- 400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227, United States
Description
XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.
Model or Catalog No. MUJ; Software Version XiO Versions 4.80 and higher
Worldwide Distribution - US Distribution including the states of IL, IN, LA, NJ, TX, WA and WI., and the countries of Belgium, Germany, Italy and Netherlands.