FDA Enforcement
Class II
Ongoing
VERIFY Dual Species Self-Contained Biological Indicators, 100 per box Item Number: S3061 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes
Recall: Z-0779-2023
·
Reported January 11, 2023
Enforcement
- Recall Number
- Z-0779-2023
- Event ID
- 91315
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Steris Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 11, 2023
- Initiation Date
- December 12, 2022
- Classification Date
- January 3, 2023
- Address
- 5960 Heisley Rd, N/A, Mentor, OH, 44060-1834, United States
Description
VERIFY Dual Species Self-Contained Biological Indicators, 100 per box Item Number: S3061 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes
Reason
Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.
Code Info
UDI: 10724995023833, 50724995023824 lot #230613
Distribution
Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA
Quantity
857 boxes