Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 20mm x10mm (OUS) UPN: M00553560 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
Enforcement
- Recall Number
- Z-0778-2021
- Event ID
- 86909
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 13, 2021
- Initiation Date
- December 1, 2020
- Classification Date
- January 6, 2021
- Termination Date
- July 14, 2023
- Address
- 300 Boston Scientific Way, Marlborough, MA, 01752-1291, United States
Description
Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 20mm x10mm (OUS) UPN: M00553560 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity
Lot Number: 26167277, 26171099, 26152171, 26171383, 26171380, 26152172 Exp. Date Range: 10/8/2021 10/12/2021 GTIN: 08714729951100
US Nationwide distribution - Foreign: TBD.
78 units