PREMIER CARE RECLINER 5570/5574 WITH CAL TB 133 UPGRADE PURCHASED JAN 01, 2011 TO AUGUST 20, 2012. A multi position recliner with a steel frame used primarily in clinics, hospitals, and treatment centers for recovery, dialysis, oncology, examination, infusion.
Enforcement
- Recall Number
- Z-0778-2013
- Event ID
- 63651
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Winco Mfg., LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- February 13, 2013
- Initiation Date
- September 24, 2012
- Classification Date
- February 4, 2013
- Termination Date
- May 23, 2013
- Address
- 5516 SW 1st Ln, N/A, Ocala, FL, 34474-9307, United States
Description
PREMIER CARE RECLINER 5570/5574 WITH CAL TB 133 UPGRADE PURCHASED JAN 01, 2011 TO AUGUST 20, 2012. A multi position recliner with a steel frame used primarily in clinics, hospitals, and treatment centers for recovery, dialysis, oncology, examination, infusion.
The performance of the chemicals in the foam and vinyl are not consistent enough to meet the TB 133 fire protection mandate required by California and 4 other states.
Model 5570 and 5574. Serial numbers: 557A100080, 557A100081, 557100004, 557100008, 557100009, 557100012, 557100041, 557100042, 557100043, 557100044, 557100045, 557A100065, 557100036, 557100037, 557100038, 557100039, 557A100073, and 557100048
Nationwide Distribution including CA, FL, IL, MA, MD, MN, NM, NY, OH, OK, and TX.
95 units (All models)