NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
Enforcement
- Recall Number
- Z-0773-2019
- Event ID
- 81920
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- NEUSOFT MEDICAL SYSTEMS IMP & EX
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- February 13, 2019
- Initiation Date
- January 9, 2019
- Classification Date
- February 5, 2019
- Termination Date
- August 25, 2022
- Address
- 16 Shiji Road, Dongling, Shenyang, N/A, China
Description
NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.
Series number includes: N1281180242E, N1281170231E, N1281180248E
US Distribution in states of: KS, OK and MA.