FDA Enforcement Class II Terminated

SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus

Recall: Z-0771-2021 · Reported January 20, 2021

Enforcement

Recall Number
Z-0771-2021
Event ID
87079
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
FDA Mandated
Initial Notification
Letter
Report Date
January 20, 2021
Initiation Date
November 24, 2020
Classification Date
January 9, 2021
Termination Date
February 6, 2024
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355, United States

Description

SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus

Reason

System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan

Code Info

SOMATOM Force (Model #10742326) SOMATOM Definition As (Model #8098027) SOMATOM Definition Edge (Model #10590000) SOMATOM Definition Flash (Model #10430603) SOMATOM Drive (Model #10431700) SOMATOM Confidence (Model #10590100) SOMATOM Edge Plus (Model #10267000) Siemens SOMATOM CT scanner systems with software syngo.CT VB20A_SP2.

Distribution

US Nationwide

Quantity

3044