Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatal Hyperbilirubinemia. 006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007296 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Enforcement
- Recall Number
- Z-0770-2017
- Event ID
- 74587
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Natus Medical Incorporated
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 21, 2016
- Initiation Date
- October 28, 2016
- Classification Date
- December 15, 2016
- Termination Date
- January 15, 2021
- Address
- 5900 1st Ave S, Seattle, WA, 98108-3248, United States
Description
Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatal Hyperbilirubinemia. 006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007296 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
neoBLUE blanket LED Phototherapy System distributed since release of the product in 2011 is recalled due to customer complaints regarding discoloration/ degradation of fiberoptic bundle at connection of pad and box.
************SERIAL NUMBERS OF UNITS DISTRIBUTED TO US GOVERMENT ACCOUNTS 000089, 000104, 000128, 001342, 001364, 001367, 001371, 001372, 001373, 001374, 001375, 001376, 001377, 001391, 001392, 001393, 001394, 001441, 001442, 001455, 001456, 001463, 001495, 001496, 001497, 001878, 001893, 001912, 001915, 001992, 001993, 002040, 002041, 002165, 002166, 002167, 002168, 002289, 002290, 002291, 002292, 002320, 002462, 002463, 002762, 002763, 002766, 002860, 002870, 002873, 002888, 003594, 003595, 003600, 003604, 003632, 003633, and 003893. ***********SERIAL NUMBERS OF UNITS DISTRIBUTED IN THE US ..000284, ..000287, ..000297, ..000299, ..000310, ..000324, ..000328, ..000365, .000301, .000313, .000321, .000322, .000323, .000352, .000415, .000419, .000420, .000426, .000427, .000435, .000445, .000449, .000467, .000483, .000508, .000518, .000523, .000531, .000541, .000581, .000651, .000670, .000695, .000696, .000697, .000698, 000030, 000031, 000032, 000033, 000035, 000036, 000039, 000041, 000042, 000043, 000044, 000046, 000048, 000050, 000051, 000052, 000053, 000056, 000062, 000070, 000076, 000081, 000082, 000085, 000086, 000087, 000088, 000093, 000094, 000139, 000140, 000143, 000144, 000146, 000162, 000163, 000164, 000166, 000167, 000170, 000180, 000181, 000187, 000189, 000190, 000193, 000195, 000202, 000205, 000208, 000215, 000218, 000221, 000223, 000234, 000235, 000238, 000240, 000242, 000243, 000244, 000245, 000246, 000248, 000249, 000250, 000253, 000254, 000255, 000257, 000263, 000275, 000276, 000281, 000290, 000291, 000292, 000293, 000303, 000317, 000320, 000325, 000332, 000333, 000335, 000336, 000340, 000343, 000344, 000345, 000349, 000350, 000351, 000352, 000353, 000354, 000366, 000411, 000414, 000417, 000418, 000422, 000424, 000425, 000428, 000430, 000432, 000436, 000441, 000442, 000443, 000446, 000452, 000453, 000454, 000456, 000457, 000458, 000459, 000460, 000462, 000469, 000472, 000475, 000477, 000479, 000483, 000489, 000493, 000495, 000499, 000500, 000502, 000511, 000516, 000522, 000528, 000530, 000547, 000551, 000553, 000560, 000561, 000563, 000565, 000570, 000574, 000599, 000604, 000615, 000616, 000619, 000620, 000625, 000628, 000631, 000641, 000643, 000646, 000647, 000648, 000655, 000656, 000662, 000664, 000667, 000671, 000675, 000680, 000681, 000682, 000683, 000684, 000687, 000693, 000694, 000697, 000708, 000709, 000711, 000712, 000715, 000717, 000718, 000719, 000720, 000731, 000732, 000734, 000735, 000736, 000740, 000741, 000743, 000744, 000745, 000749, 000751, 000752, 000754, 000758, 000760, 000762, 000763, 000766, 000767, 000768, 000769, 000770, 000771, 000772, 000774, 000775, 000776, 000781, 000782, 000783, 000784, 000785, 000786, 000787, 000788, 000789, 000790, 000792, 000793, 000794, 000796, 000797, 000798, 000799, 000800, 000801, 000802, 000803, 000804, 000805, 000806, 000807, 000808, 000809, 000810, 000812, 000813, 000814, 000815, 000816, 000817, 000818, 000820, 000821, 000822, 000825, 000826, 000831, 000833, 000836, 000847, 000848, 000849, 000850, 000851, 000852, 000854, 000857, 000858, 000859, 000861, 000862, 000863, 000865, 000866, 000867, 000868, 000869, 000870, 000871, 000873, 000874, 000875, 000876, 000877, 000878, 000879, 000880, 000882, 000884, 000891, 000892, 000893, 000899, 000900, 000901, 000902, 000903, 000904, 000905, 000907, 000908, 000909, 000910, 000912, 000913, 000914, 000915, 000916, 000921, 000922, 000923, 000924, 000925, 000935, 000936, 000937, 000938, 000939, 000940, 000941, 000943, 000944, 000945, 000946, 000947, 000950, 000952, 000953, 000957, 000958, 000959, 000960, 000961, 000962, 000963, 000965, 000966, 000967, 000970, 000972, 000973, 000974, 000976, 000978, 000979, 000981, 000988, 000991, 000992, 000993, 001005, 001006, 001010, 001011, 001012, 001013, 001018, 001019, 001020, 001021, 001022, 001031, 001035, 001036, 001037, 001038, 001039, 001040, 001041, 001042, 001043, 001044, 001046, 001047, 001048, 001050, 001056, 001058, 001059, 001085, 001086, 001087, 001088, 001095, 001103, 001104, 001105, 001106, 001109, 001110, 001111, 001113, 001115, 001117, 001118, 001119, 001121, 001122, 001123, 001124, 001125, 001126, 001127, 001128, 001129, 001130, 001131, 001133, 001134, 001135, 001136, 001137, 001138, 001139, 001140, 001141, 001143, 001144, 001146, 001149, 001150, 001153, 001154, 001155, 001156, 001161, 001162, 001163, 001164, 001165, 001166, 001167, 001190, 001205, 001206, 001207, 001208, 001209, 001211, 001213, 001214, 001215, 001216, 001217, 001218, 001219, 001220, 001221, 001222, 001223, 001224, 001225, 001227, 001228, 001230, 001231, 001233, 001235, 001236, 001239, 001241, 001242, 001249, 001252, 001253, 001254, 001255, 001256, 001257, 001260, 001261, 001262, 001265, 001266, 001268, 001269, 001271, 001272, 001274, 001275, 001277, 001280, 001281, 001283, 001288, 001289, 001290, 001291, 001292, 001293, 001294, 001296, 001297, 001299, 001300, 001301, 001302, 001305, 001307, 001309, 001311, 001312, 001315, 001316, 001319, 001321, 001327, 001329, 001330, 001331, 001334, 001335, 001338, 001340, 001346, 001348, 001349, 001350, 001351, 001353, 001354, 001356, 001357, 001360, 001361, 001362, 001363, 001366, 001368, 001370, 001378, 001379, 001380, 001381, 001382, 001383, 001384, 001385, 001386, 001387, 001388, 001389, 001390, 001397, 001398, 001399, 001400, 001401, 001402, 001403, 001404, 001406, 001407, 001408, 001409, 001411, 001414, 001415, 001416, 001419, 001420, 001425, 001427, 001428, 001429, 001430, 001431, 001432, 001433, 001434, 001435, 001436, 001437, 001439, 001440, 001443, 001444, 001445, 001446, 001447, 001452, 001454, 001457, 001467, 001470, 001472, 001473, 001474, 001475, 001476, 001477, 001478, 001481, 001486, 001492, 001494, 001499, 001500, 001503, 001507, 001528, 001532, 001540, 001541, 001542, 001543, 001544, 001545, 001546, 001547, 001548, 001549, 001550, 001551, 001552, 001553, 001554, 001556, 001557, 001558, 001559, 001560, 001562, 001564, 001565, 001566, 001567, 001568, 001569, 001579, 001580, 001581, 001582, 001583, 001584, 001586, 001587, 001588, 001589, 001590, 001591, 001592, 001593, 001594, 001596, 001597, 001598, 001599, 001600, 001601, 001602, 001603, 001607, 001613, 001614, 001615, 001616, 001617, 001618, 001619, 001620, 001622, 001623, 001624, 001625, 001626, 001627, 001629, 001630, 001631, 001632, 001633, 001634, 001636, 001637, 001638, 001640, 001644, 001645, 001646, 001647, 001648, 001649, 001650, 001651, 001652, 001656, 001657, 001658, 001661, 001662, 001663, 001665, 001666, 001667, 001680, 001681, 001682, 001683, 001685, 001686, 001687, 001689, 001691, 001693, 001695, 001696, 001698, 001701, 001702, 001703, 001709, 001711, 001712, 001714, 001715, 001716, 001717, 001718, 001719, 001720, 001721, 001729, 001733, 001738, 001740, 001741, 001742, 001748, 001752, 001753, 001755, 001757, 001758, 001760, 001761, 001762, 001764, 001765, 001766, 001769, 001771, 001772, 001774, 001777, 001779, 001780, 001781, 001785, 001799, 001813, 001814, 001815, 001817, 001818, 001821, 001822, 001823, 001834, 001836, 001840, 001844, 001845, 001846, 001848, 001850, 001851, 001852, 001855, 001856, 001858, 001860, 001861, 001862, 001863, 001864, 001866, 001867, 001868, 001869, 001872, 001873, 001874, 001875, 001876, 001879, 001880, 001881, 001882, 001883, 001890, 001894, 001895, 001896, 001897, 001898, 001899, 001900, 001902, 001903, 001904, 001905, 001906, 001907, 001908, 001909, 001910, 001916, 001917, 001918, 001919, 001920, 001921, 001922, 001923, 001924, 001925, 001926, 001927, 001928, 001929, 001930, 001932, 001935, 001936, 001937, 001938, 001939, 001940, 001941, 001942, 001943, 001946, 001947, 001948, 001949, 001950, 001951, 001953, 001954, 001955, 001956, 001957, 001958, 001959, 001960, 001961, 001962, 001963, 001964, 001965, 001967, 001969, 001970, 001973, 001974, 001979, 001982, 001983, 001984, 001987, 001988, 001990, 001991, 001994, 002001, 002009, 002010, 002014, 002016, 002017, 002018, 002019, 002020, 002021, 002022, 002023, 002024, 002026, 002027, 002031, 002032, 002034, 002037, 002038, 002043, 002044, 002045, 002046, 002047, 002048, 002049, 002075, 002076, 002077, 002078, 002079, 002080, 002086, 002087, 002088, 002089, 002090, 002091, 002092, 002141, 002142, 002143, 002144, 002145, 002146, 002147, 002148, 002150, 002151, 002152, 002154, 002156, 002157, 002158, 002159, 002160, 002161, 002162, 002163, 002164, 002169, 002170, 002171, 002172, 002173, 002175, 002176, 002179, 002180, 002181, 002182, 002183, 002185, 002186, 002198, 002199, 002200, 002201, 002202, 002203, 002208, 002209, 002212, 002215, 002217, 002218, 002219, 002225, 002226, 002227, 002228, 002233, 002234, 002235, 002236, 002237, 002239, 002240, 002241, 002243, 002247, 002248, 002249, 002250, 002252, 002253, 002254, 002259, 002260, 002261, 002262, 002263, 002264, 002266, 002267, 002268, 002269, 002270, 002271, 002272, 002273, 002274, 002275, 002276, 002277, 002281, 002284, 002285, 002286, 002287, 002288, 002293, 002294, 002295, 002296, 002305, 002306, 002308, 002309, 002310, 002311, 002312, 002313, 002315, 002316, 002318, 002319, 002320, 002321, 002322, 002323, 002324, 002325, 002333, 002334, 002335, 002337, 002338, 002339, 002344, 002345, 002352, 002356, 002357, 002359, 002360, 002361, 002363, 002366, 002367, 002368, 002369, 002370, 002372, 002376, 002378, 002379, 002380, 002381, 002382, 002383, 002386, 002387, 002388, 002389, 002392, 002393, 002394, 002395, 002396, 002397, 002398, 002399, 002400, 002401, 002403, 002404, 002409, 002415, 002416, 002417, 002421, 002422, 002429, 002430, 002431, 002432, 002435, 002436, 002437, 002438, 002439, 002440, 002441, 002442, 002443, 002444, 002445, 002446, 002448, 002449, 002451, 002452, 002453, 002454, 002455, 002456, 002457, 002458, 002459, 002460, 002461, 002469, 002471, 002473, 002476, 002478, 002480, 002482, 002483, 002484, 002485, 002486, 002488, 002489, 002490, 002491, 002492, 002493, 002494, 002495, 002496, 002497, 002498, 002499, 002501, 002502, 002503, 002504, 002505, 002506, 002507, 002508, 002509, 002513, 002514, 002516, 002517, 002519, 002520, 002521, 002527, 002528, 002529, 002530, 002533, 002536, 002539, 002543, 002544, 002545, 002546, 002652, 002653, 002654, 002655, 002657, 002658, 002659, 002660, 002661, 002662, 002663, 002669, 002670, 002671, 002672, 002673, 002675, 002676, 002694, 002695, 002696, 002697, 002698, 002699, 002700, 002701, 002702, 002703, 002704, 002715, 002718, 002719, 002722, 002723, 002724, 002727, 002728, 002729, 002732, 002733, 002735, 002736, 002738, 002739, 002751, 002752, 002753, 002754, 002755, 002757, 002758, 002765, 002767, 002768, 002770, 002771, 002773, 002774, 002775, 002776, 002777, 002778, 002779, 002781, 002782, 002783, 002786, 002787, 002790, 002791, 002792, 002793, 002799, 002800, 002801, 002803, 002807, 002808, 002810, 002811, 002812, 002813, 002814, 002815, 002816, 002817, 002818, 002819, 002820, 002821, 002822, 002823, 002824, 002825, 002827, 002828, 002829, 002830, 002831, 002836, 002837, 002838, 002839, 002840, 002841, 002842, 002858, 002859, 002861, 002863, 002866, 002867, 002868, 002869, 002871, 002872, 002874, 002875, 002876, 002877, 002878, 002879, 002880, 002884, 002886, 002887, 002889, 002890, 002891, 002892, 002893, 002894, 002895, 002899, 002910, 002911, 002913, 002914, 002919, 002926, 002931, 002934, 002946, 002947, 002952, 002959, 002960, 002961, 002962, 002963, 002964, 002965, 002966, 002967, 002968, 002969, 002970, 002971, 002972, 002973, 002974, 002975, 002976, 002977, 002979, 002980, 002981, 002982, 002983, 002984, 002985, 002986, 002987, 002988, 002989, 002990, 002991, 002992, 002993, 002994, 003007, 003009, 003010, 003014, 003015, 003016, 003018, 003019, 003020, 003021, 003022, 003023, 003025, 003026, 003027, 003028, 003029, 003030, 003031, 003032, 003033, 003034, 003035, 003036, 003037, 003039, 003040, 003041, 003042, 003043, 003044, 003045, 003046, 003047, 003053, 003068, 003069, 003070, 003071, 003072, 003074, 003075, 003076, 003077, 003078, 003083, 003095, 003102, 003103, 003104, 003105, 003106, 003107, 003108, 003111, 003112, 003113, 003114, 003115, 003116, 003117, 003118, 003119, 003120, 003121, 003122, 003123, 003124, 003125, 003127, 003128, 003129, 003135, 003136, 003137, 003138, 003139, 003140, 003141, 003142, 003143, 003149, 003150, 003151, 003152, 003153, 003154, 003155, 003156, 003160, 003161, 003583, 003584, 003590, 003592, 003593, 003596, 003597, 003601, 003603, 003605, 003606, 003607, 003608, 003609, 003610, 003611, 003613, 003614, 003615, 003616, 003617, 003618, 003619, 003620, 003621, 003622, 003623, 003624, 003625, 003626, 003627, 003629, 003630, 003631, 003634, 003638, 003639, 003640, 003641, 003642, 003643, 003645, 003646, 003647, 003648, 003649, 003650, 003651, 003657, 003658, 003667, 003668, 003669, 003670, 003671, 003672, 003673, 003674, 003675, 003680, 003681, 003682, 003683, 003684, 003685, 003686, 003687, 003688, 003690, 003691, 003692, 003693, 003694, 003695, 003696, 003697, 003698, 003699, 003700, 003701, 003702, 003708, 003709, 003710, 003711, 003712, 003713, 003714, 003715, 003716, 003717, 003718, 003719, 003720, 003721, 003722, 003723, 003724, 003728, 003729, 003730, 003731, 003732, 003733, 003736, 003737, 003739, 003740, 003743, 003744, 003745, 003746, 003747, 003749, 003750, 003751, 003752, 003753, 003754, 003755, 003756, 003757, 003759, 003760, 003761, 003762, 003763, 003765, 003766, 003767, 003772, 003773, 003774, 003775, 003780, 003786, 003788, 003789, 003791, 003792, 003793, 003794, 003795, 003796, 003797, 003798, 003799, 003800, 003801, 003802, 003803, 003804, 003805, 003806, 003809, 003810, 003811, 003812, 003813, 003814, 003815, 003816, 003817, 003821, 003822, 003823, 003832, 003833, 003834, 003836, 003837, 003838, 003840, 003841, 003845, 003846, 003847, 003853, 003855, 003856, 003858, 003859, 003860, 003861, 003862, 003863, 003866, 003868, 003869, 003870, 003871, 003872, 003873, 003875, 003877, 003878, 003879, 003882, 003885, 003891, 003892, 003896, 003898, 003899, 003900, 003904, 003908, 003910, 003911, 003912, 003913, 003915, 003916, 003918, 003919, 003921, 003924, 003926, 003927, 003930, 003932, 003933, 003934, 003935, 003936, 003937, 003938, 003939, 003940, 003941, 003942, 003943, 003944, 003945, 003946, 003947, 003948, 003949, 003950, 003951, 003952, 003953, 003954, 003955, 003957, 003959, 003961, 003962, 003963, 003964, 003965, 003966, 003967, 003968, 003970, 003971, 003972, 003975, 003976, 003977, 003978, 003979, 003980, 003981, 003982, 003983, 003984, 003988, 003989, 003990, 003991, 003992, 003994, 003995, 003996, 003997, 003998, 004000, 004001, 004002, 004003, 004004, 004005, 004006, 004007, 004008, 004010, 004011, 004012, 004013, 004014, 004015, 004016, 004017, 004018, 004019, 004020, 004021, 004022, 004023, 004024, 004026, 004027, 004028, 004029, 004032, 004033, 004034, 004035, 004041, 004042, 004043, 004072, 004207, 004208, 004209, 004210, 004211, 004212, 004213, 004214, 004216, 004217, 004218, 004219, 004220, 004221, 004222, 004223, 004224, 004225, 004226, 004227, 004228, 004229, 004231, 004232, 004233, 004234, 004236, 004237, 004238, 004240, 004241, 004246, 004247, 004248, 004249, 004250, 004251, 004252, 004253, 004254, 004255, 004256, 004257, 004258, 004259, 004260, 004261, 004263, 004264, 004268, 004269, 004270, 004271, 004273, 1215R002299, 2298, 2300, 2302, 2303, 2304, NR002392, R002300, R002301, R002357, R002359, R002360, R002362, R002364, R002365, R002368, R002369, R002370, R002389, R002413, R002415, R002416, and R002417. ************SERIAL NUMBERS OF UNITS DISTRIBUTED OUTSIDE US ..000283, 000061, 000064, 000066, 000067, 000069, 000078, 000079, 000121, 000123, 000125, 000129, 000130, 000131, 000136, 000137, 000147, 000148, 000149, 000150, 000152, 000153, 000155, 000168, 000169, 000173, 000174, 000182, 000183, 000192, 000194, 000196, 000198, 000200, 000201, 000211, 000212, 000216, 000219, 000227, 000280, 000295, 000296, 000298, 000300, 000302, 000306, 000309, 000315, 000326, 000334, 000356, 000357, 000368, 000369, 000412, 000423, 000429, 000437, 000451, 000471, 000515, 000521, 000536, 000537, 000539, 000542, 000544, 000548, 000549, 000550, 000581, 000586, 000589, 000590, 000591, 000592, 000593, 000607, 000608, 000609, 000610, 000617, 000624, 000626, 000630, 000633, 000635, 000636, 000637, 000638, 000639, 000642, 000650, 000653, 000654, 000657, 000672, 000673, 000685, 000686, 000691, 000692, 000701, 000706, 000707, 000713, 000714, 000716, 000721, 000723, 000724, 000725, 000726, 000728, 000729, 000737, 000738, 000739, 000746, 000747, 000748, 000757, 000764, 000765, 000777, 000778, 000779, 000780, 000840, 000885, 000886, 000887, 000888, 000889, 000890, 000896, 000897, 000918, 000919, 000920, 000931, 000932, 000933, 000934, 000956, 000975, 000977, 000980, 000994, 000995, 000996, 000997, 001001, 001002, 001007, 001008, 001009, 001014, 001015, 001016, 001017, 001049, 001051, 001052, 001053, 001054, 001055, 001057, 001060, 001061, 001062, 001063, 001064, 001065, 001066, 001069, 001070, 001071, 001072, 001073, 001074, 001075, 001076, 001077, 001078, 001079, 001080, 001081, 001082, 001083, 001084, 001089, 001090, 001091, 001092, 001093, 001097, 001098, 001099, 001100, 001101, 001102, 001114, 001172, 001178, 001179, 001187, 001188, 001192, 001193, 001194, 001198, 001244, 001245, 001246, 001247, 001248, 001250, 001251, 001320, 001325, 001326, 001328, 001332, 001333, 001336, 001339, 001341, 001343, 001345, 001352, 001355, 001358, 001359, 001365, 001395, 001396, 001417, 001418, 001423, 001450, 001451, 001453, 001458, 001459, 001461, 001462, 001479, 001490, 001491, 001509, 001511, 001512, 001513, 001514, 001515, 001516, 001519, 001520, 001521, 001524, 001526, 001561, 001570, 001571, 001572, 001573, 001574, 001576, 001577, 001578, 001604, 001605, 001606, 001608, 001642, 001643, 001659, 001660, 001668, 001669, 001671, 001672, 001673, 001674, 001675, 001676, 001677, 001678, 001694, 001726, 001727, 001728, 001730, 001731, 001732, 001734, 001735, 001736, 001743, 001744, 001745, 001746, 001749, 001750, 001751, 001763, 001773, 001789, 001793, 001796, 001798, 001801, 001802, 001804, 001843, 001859, 001884, 001885, 001888, 001889, 001891, 001892, 001933, 001934, 001968, 001976, 001977, 001985, 001995, 001996, 001997, 001998, 001999, 002000, 002002, 002003, 002004, 002005, 002007, 002011, 002012, 002013, 002025, 002028, 002036, 002039, 002051, 002052, 002053, 002054, 002055, 002056, 002057, 002058, 002059, 002061, 002062, 002065, 002066, 002067, 002068, 002069, 002070, 002071, 002072, 002073, 002074, 002081, 002082, 002083, 002084, 002085, 002187, 002188, 002191, 002192, 002193, 002194, 002195, 002197, 002204, 002205, 002206, 002207, 002210, 002211, 002213, 002214, 002216, 002220, 002221, 002222, 002223, 002224, 002229, 002230, 002231, 002232, 002244, 002245, 002255, 002256, 002257, 002282, 002307, 002318, 002342, 002347, 002350, 002351, 002353, 002354, 002364, 002371, 002372, 002373, 002374, 002375, 002376, 002382, 002383, 002384, 002385, 002391, 002394, 002401, 002402, 002404, 002405, 002406, 002407, 002408, 002409, 002410, 002411, 002412, 002413, 002414, 002423, 002424, 002425, 002426, 002428, 002450, 002464, 002465, 002467, 002477, 002479, 002510, 002511, 002531, 002532, 002534, 002535, 002664, 002665, 002666, 002667, 002668, 002677, 002679, 002681, 002682, 002683, 002684, 002687, 002688, 002689, 002691, 002692, 002693, 002706, 002708, 002709, 002710, 002712, 002713, 002714, 002720, 002730, 002731, 002734, 002737, 002740, 002741, 002742, 002743, 002744, 002745, 002746, 002747, 002748, 002749, 002750, 002756, 002759, 002760, 002761, 002764, 002772, 002797, 002804, 002805, 002806, 002809, 002826, 002834, 002835, 002864, 002865, 002881, 002882, 002916, 002920, 002923, 002927, 002928, 002929, 002930, 002932, 002933, 002935, 002936, 002937, 002938, 002939, 002940, 002941, 002942, 002943, 002944, 002945, 002948, 002949, 002950, 002954, 002955, 002956, 002957, 002996, 002997, 002999, 003000, 003001, 003002, 003003, 003005, 003011, 003012, 003017, 003038, 003051, 003052, 003055, 003056, 003057, 003058, 003059, 003060, 003061, 003062, 003063, 003065, 003066, 003067, 003073, 003081, 003084, 003085, 003086, 003087, 003088, 003089, 003090, 003091, 003092, 003093, 003094, 003096, 003097, 003098, 003099, 003100, 003101, 003130, 003131, 003132, 003133, 003134, 003144, 003145, 003146, 003148, 003157, 003158, 003159, 003162, 003585, 003586, 003587, 003588, 003591, 003598, 003599, 003636, 003637, 003644, 003652, 003653, 003654, 003655, 003656, 003676, 003678, 003704, 003705, 003706, 003707, 003725, 003726, 003727, 003734, 003735, 003738, 003741, 003742, 003770, 003771, 003776, 003777, 003778, 003779, 003781, 003782, 003790, 003807, 003808, 003818, 003819, 003820, 003835, 003842, 003843, 003844, 003848, 003849, 003850, 003851, 003852, 003854, 003864, 003865, 003867, 003876, 003880, 003881, 003884, 003886, 003887, 003888, 003890, 003895, 003897, 003902, 003903, 003905, 003906, 003907, 003909, 003917, 003923, 003925, 003931, 003960, 003969, 003993, 004242, 004243, 004267, 004272, 1517, 1911000001, NR002393, R002358, R002366, R002371, R002390, and R002414.
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : including American Samoa and Guam, and in the following countries: Albania, Argentina, Australia, Austria, Bangladesh, Canada, Chile, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Hong Kong, Indonesia, Iraq, Israel, Italy, Korea, Kuwait, Latvia, Malaysia, Mongolia, Namibia, Netherlands, Nigeria, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia (Slovak Republic), Slovenia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Virgin Islands (British).
2619 units (1925 units in the US and 694 units OUTSIDE the US)