Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS

Recall: Z-0769-2017 · Reported December 21, 2016

Enforcement

Recall Number
Z-0769-2017
Event ID
75228
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Natus Medical Incorporated
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 21, 2016
Initiation Date
September 16, 2016
Classification Date
December 15, 2016
Termination Date
September 1, 2017
Address
5900 1st Ave S, Seattle, WA, 98108-3248, United States

Description

Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS

Reason

The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.

Code Info

*************Serial Numbers of Units distributed in the US ************** 004329, 004330, 004349, 004360, 004361, 004362, 004363, 004364, 004365, 004366, 004367, 004369, 004370, 004371, 004417, 004421, 004426, 004436, 004437, 004443, 004448, 004449, 004454, 004456, 004473, 004481, 004488, 004490, 004494, 004495, 004502, 004507, 004517, 004523, 004525, 004526, 004527, 004528, 004530, 004533, 004539, 004541, 004544, 004545, 004546, 004548, 004552, 004553, 004554, 004555, 004556, 004557, 004558, 004567, 004568, 004573, 004578, 004581, 004582, 004584, 004588, 004591, 004602, 004608, 004630, 004632, 004637, 004640, 004647, 004659, 004667, 004668, 004670, 004672, 004673, 004677, 004678, 004679, 004681, 004683, 004684, 004685, 004686, 004693, 004694, 004696, 004700, 004706, 004707, 004708, 004709, 004710, 004711, 004712, 004713, 004714, 004715, 004716, 004717, 004718, 004719, 004720, 004721, 004722, 004723, 004724, 004725, 004726, 004727, 004728, 004729, 004730, 004731, 004732, 004733, 004734, 004735, 004736, 004737, 004738, 004739, 004740, 004741, 004742, 004743, 004745, 004746, 004747, 004749, 004750, 004753, 004754, 004756, 004757, 004760, 004761, 004762, 004763, 004764, 004766, 004767, 004768, 004771, 004772, 004773, 004774, 004775, 004776, 004777, 004778, 004779, 004780, 004781, 004782, 004783, 004784, 004785, 004787, 004788, 004789, 004790, 004791, 004793, 004794, 004795, 004796, 004797, 004798, 004800, 004801, 004802, 004803, 004804, 004805, 004806, 004807, 004808, 004810, 004811, 004812, 004814, 004815, 004816, 004818, 004819, 004820, 004822, 004823, 004824, 004825, 004826, 004827, 004828, 004829, 004830, 004831, 004832, 004833, 004834, 004835, 004836, 004837, 004838, 004839, 004840, 004841, 004842, 004843, 004844, 004845, 004846, 004847, 004848, 004849, 004850, 004851, 004852, 004853, 004854, 004855, 004856, 004857, 004858, 004859, 004860, 004861, 004862, 004863, 004864, 004865, 004866, 004867, 004868, 004869, 004870, 004871, 004872, 004873, 004874, 004876, 004877, 004878, 004879, 004880, 004881, 004882, 004884, 004885, 004886, 004887, 004888, 004890, 004891, 004893, 004894, 004895, 004896, 004897, 004899, 004900, 004901, 004902, 004903, 004904, 004905, 004906, 004907, 004909, 004910, 004997, 004998, 004999, 005007, 005008, 005009, 005010, 005011, 005012, 005013, 005014, 005015, 005016, 005017, 005018, 005019, 005020, 005021, 005022, 005023, 005024, 005025, 005026, 005027, 005028, 005029, 005030, 005031, 005032, 005033, 005034, 005035, 005036, 005037, 005039, 005040, 005041, 005042, 005043, 005044, 005045, 005046, 005047, 005048, 005049, 005050, 005051, 005052, 005053, 005054, 005055, 005056, 005057, 005058, 005059, 005060, 005061, 005062, 005063, 005064, 005065, 005066, 005067, 005068, 005069, 005070, 005071, 005072, 005073, 005074, 005075, 005076, 005077, 005078, 005079, 005080, 005081, 005082, 005083, 005084, 005085, 005086, 005087, 005088, 005089, 005090, 005091, 005092, 005093, 005094, 005095, 005096, 005097, 005098, 005099, 005100, 005101, 005103, 005104, 005105, 005106, 005107, 005108, 005112, 005113, 005115, 005117, 005118, 005119, 005120, 005121, 005122, 005123, 005124, 005125, 005128, 005136, 005137, 005138, 005139, 005140, 005141, 005142, 005143, 005144, 005145, 005146, 005147, 005148, 005149, 005150, 005151, 005152, and 005153. *************Serial Numbers of Units distributed in the US ************** 004283, 004284, 004285, 004286, 004287, 004288, 004290, 004291, 004292, 004293, 004294, 004295, 004296, 004298, 004299, 004302, 004303, 004305, 004306, 004307, 004308, 004309, 004310, 004312, 004313, 004314, 004315, 004316, 004319, 004320, 004321, 004322, 004325, 004326, 004327, 004328, 004332, 004333, 004334, 004335, 004337, 004338, 004339, 004340, 004341, 004343, 004344, 004345, 004347, 004348, 004350, 004351, 004353, 004354, 004355, 004356, 004357, 004359, 004372, 004373, 004374, 004375, 004379, 004380, 004381, 004382, 004383, 004384, 004385, 004386, 004388, 004389, 004390, 004391, 004392, 004393, 004394, 004395, 004396, 004397, 004399, 004400, 004401, 004402, 004403, 004404, 004405, 004406, 004408, 004411, 004412, 004413, 004414, 004415, 004416, 004418, 004419, 004420, 004422, 004423, 004424, 004425, 004427, 004428, 004430, 004431, 004432, 004433, 004439, 004440, 004442, 004445, 004446, 004447, 004450, 004451, 004452, 004453, 004455, 004457, 004458, 004459, 004460, 004462, 004463, 004464, 004465, 004466, 004467, 004468, 004469, 004471, 004472, 004474, 004475, 004476, 004477, 004478, 004479, 004480, 004484, 004485, 004486, 004489, 004491, 004492, 004496, 004497, 004498, 004499, 004500, 004501, 004503, 004504, 004505, 004508, 004510, 004511, 004512, 004513, 004514, 004515, 004516, 004518, 004519, 004520, 004521, 004522, 004524, 004531, 004532, 004536, 004542, 004543, 004549, 004551, 004565, 004569, 004570, 004571, 004572, 004574, 004575, 004576, 004577, 004585, 004587, 004589, 004590, 004592, 004593, 004594, 004596, 004597, 004598, 004599, 004600, 004601, 004604, 004605, 004606, 004609, 004610, 004611, 004612, 004613, 004614, 004615, 004616, 004617, 004618, 004619, 004620, 004621, 004622, 004623, 004624, 004625, 004626, 004628, 004629, 004631, 004633, 004634, 004635, 004638, 004639, 004641, 004642, 004643, 004644, 004645, 004646, 004648, 004649, 004650, 004651, 004652, 004653, 004655, 004656, 004657, 004658, 004660, 004662, 004663, 004664, 004665, 004666, 004669, 004671, 004674, 004675, 004680, 004682, 004688, 004689, 004690, 004692, 004695, 004697, 004698, 004699, 004701, 004702, 004703, 004704, 004705, 004744, 004748, 004751, 004752, 004755, 004758, 004765, 004769, 004770, 004792, 004799, 004809, 004813, 004817, 004821, 004875, 004889, 004898, 004908, 005102, 005111, 005114, 005116, 005126, 005127, 005129, 005130, 005131, 005132, 005133, 005134, 005135, 005154, 005155, 005239, 005240, 005241, 4317, 4323, 4324, 4368, and 4377.

Distribution

Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, NC, ND, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, WV and Puerto Rico. , and to the countries of : Argentina, Austria, Canada, Chile, China, Denmark, Dominican Republic, Finland, France, Germany, Hong Kong, Hungary, Indonesia, Italy, Latvia, Malaysia, Namibia, Paraguay, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Taiwan, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Virgin Islands (British).

Quantity

713 units total (403 units in US and 310 units outside US)