Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray system suitable for radiography and fluoroscopy examinations. The system is stationary mounted. The movements of the table and the tube stand are motorized and remotely controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
Enforcement
- Recall Number
- Z-0764-2020
- Event ID
- 84309
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Villa Sistemi Medicali S.P.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 8, 2020
- Initiation Date
- November 4, 2019
- Classification Date
- December 31, 2019
- Address
- Via delle Azalee 3, N/A, Buccinasco, N/A, N/A, Italy
Description
Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray system suitable for radiography and fluoroscopy examinations. The system is stationary mounted. The movements of the table and the tube stand are motorized and remotely controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
Serial Numbers: 08020473 11100873 15041446
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.