FDA Enforcement Class II Terminated

MCA7-08SD - 0.8MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Recall: Z-0763-2014 · Reported January 22, 2014

Enforcement

Recall Number
Z-0763-2014
Event ID
63624
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
The Anspach Effort, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 22, 2014
Initiation Date
March 29, 2012
Classification Date
January 14, 2014
Termination Date
March 11, 2015
Address
4500 Riverside Drive, Palm Beach Gardens, FL, 33410-4235, United States

Description

MCA7-08SD - 0.8MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Reason

The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide

Code Info

All lots are included in recall

Distribution

Worldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.

Quantity

6439 total