FDA Enforcement Class I Terminated

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

Recall: Z-0762-2019 · Reported February 6, 2019

Enforcement

Recall Number
Z-0762-2019
Event ID
81878
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Edwards Lifesciences, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 6, 2019
Initiation Date
December 21, 2018
Classification Date
January 31, 2019
Termination Date
November 2, 2023
Address
1 Edwards Way, N/A, Irvine, CA, 92614-5688, United States

Description

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

Reason

These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.

Code Info

Lot Numbers: 61321177, 61227528

Distribution

Worldwide Distribution

Quantity

277 units