FDA Enforcement Class II Terminated

The Randox Liquid Cardiac Controls Level 1, Level 2, and Level 3 are liquid controls containing BNP, CK MB Mass, Digoxin, D-Dimer, Homocysteine, hsCRP, Myoglobin, NT-ProBNP, Troponin I, and Troponin T. They have been developed for use in the quality control of BNP, CK MB Mass, Digoxin, D-Dimer, Homocysteine, hsCRP, Myoglobin, NT-ProBNP, Troponin I, and Troponin T assays on various clinical chemistry and immunoassay systems. This in vitro diagnostic device is intended for prescription use only.

Recall: Z-0762-2017 · Reported December 21, 2016

Enforcement

Recall Number
Z-0762-2017
Event ID
75262
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Randox Laboratories
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
December 21, 2016
Initiation Date
September 16, 2016
Classification Date
December 14, 2016
Termination Date
April 11, 2018
Address
55 The Diamond Road, Antrim, N/A, United Kingdom

Description

The Randox Liquid Cardiac Controls Level 1, Level 2, and Level 3 are liquid controls containing BNP, CK MB Mass, Digoxin, D-Dimer, Homocysteine, hsCRP, Myoglobin, NT-ProBNP, Troponin I, and Troponin T. They have been developed for use in the quality control of BNP, CK MB Mass, Digoxin, D-Dimer, Homocysteine, hsCRP, Myoglobin, NT-ProBNP, Troponin I, and Troponin T assays on various clinical chemistry and immunoassay systems. This in vitro diagnostic device is intended for prescription use only.

Reason

An internal real-time stability monitoring program revealed decreased recovery of myoglobin (-20% bias) for Randox Immunoturbidimetric and Siemens Centaur Classic methods and CKMB (-21% bias) for Roche e411, BioMerieux Vidas, and Siemens Centaur Classic methods.

Code Info

CQ5053 CQ5052

Distribution

Worldwide Distribution - US Distribution.

Quantity

2817