FDA Enforcement Class II Terminated

Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP31 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.

Recall: Z-0760-2019 · Reported January 30, 2019

Enforcement

Recall Number
Z-0760-2019
Event ID
81755
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stihler Electronic Gmbh
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
January 30, 2019
Initiation Date
September 6, 2018
Classification Date
January 23, 2019
Termination Date
May 13, 2020
Address
Julius-Holder-Str. 36, N/A, Stuttgart, N/A, N/A, Germany

Description

Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP31 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.

Reason

PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn

Code Info

Serial Numbers: E22368

Distribution

Worldwide Distribution - US Nationwide in the states of CA and Foreign of: Switzerland

Quantity

1 unit