FDA Enforcement Class II Ongoing

Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40 - ARDLCA209021A, LCA 40 - ARDLCA209021C, LCA 40 MOBILE - ARDLCA309009A, LCA 40 WALL - ARDLCA109001A, LCA 40GS - ARDLCA109009A, LCA 40GS - ARDLCA109009C, LCA 100 - ARDLCA109005A, LCA 100 - ARDLCA109005C, LCA 100 - ARDLCA219000A, LCA 100 MOBILE - ARDLCA309006A, LCA 100 MOBILE B - ARDLCA309007A, LCA 100 SAT - ARDSAT209004A, LCA 100 SAT - ARDSAT209005A, LCA 100 V - ARDLCA219000C, LCA 100DF - ARDLCA109008A, LCA 100DF - ARDLCA109008C, LCA 100DF - ARDLCA219003A, LCA 100DF - ARDSAT209008A, LCA 100DF - ARDSAT209009A, LCA 100DF V - ARDLCA219003C, LCA 50 - ARDLCA109004A, LCA 50 - ARDLCA109004C, LCA 50 - ARDLCA209012A, LCA 50 - ARDLCA209012C, LCA 50 - ARDSAT209002A, LCA 50 MOBILE - ARDLCA309004A, LCA 50 MOBILE B - ARDLCA309005A, LCA 50 SAT - ARDSAT209003A, LCA 50 WALL 11 - ARD2LCA00140A, LCA 50DF - ARDLCA109007A, LCA 50DF - ARDLCA109007C, LCA 50DF - ARDLCA209015A, LCA 50DF - ARDLCA209015C, LCA 50DF - ARDSAT209006A, LCA 50DF - ARDSAT209007A, LCA DUO 100 - ARDLCA219001A, LCA DUO 100 V - ARDLCA219001C, LCA DUO 100+50 - ARDLCA219002A, LCA DUO 100+50 V - ARDLCA219002C, LCA DUO 100+50DF - ARDLCA219005A, LCA DUO 100+50DF V - ARDLCA219005C, LCA DUO 100DF - ARDLCA219004A, LCA DUO 100DF V - ARDLCA219004C, LCA DUO 50 - ARDLCA209013A, LCA DUO 50 - ARDLCA209013C, LCA DUO 50DF - ARDLCA209016A, LCA DUO 50DF - ARDLCA209016C, LCA100 MOBILE UPGRADE - ARDLCA506006A, LCA100 UPGRADE - ARDLCA506004A, LCA100 UPGRADE 24V - ARDLCA506005A, LCA100 WALL UPGRADE - ARDLCA506007A, LCA50 UPGRADE - ARDLCA506000A, LCA50 UPGRADE 24V - ARDLCA506001A, LCA50 UPGRADE WALL MOUNTED - ARDLCA506003A, SAT LCA 100 - ARDSAT209004C, SAT LCA 100 - ARDSAT209005C, SAT LCA 100DF - ARDSAT209008C, SAT LCA 100DF - ARDSAT209009C, SAT LCA 50 - ARDSAT209002C, SAT LCA 50 - ARDSAT209003C, SAT LCA 50DF - ARDSAT209006C, SAT LCA 50DF - ARDSAT209007C, LCA 10 MOB - ARDLCA309000A, LCA 10 MOB - ARDLCA309000C, LCA 10 MOB - ARDLCA309002A, LCA 10 MOB - ARDLCA309002C, LCA 10 MOB - ARDLCA309008C, LCA 100 MOBILE - ARDLCA309006C, LCA 100 MOBILE B - ARDLCA309007C, LCA 40 MOB - ARDLCA309001A, LCA 40 MOB - ARDLCA309001C, LCA 40 MOB - ARDLCA309003A, LCA 40 MOB - ARDLCA309003C, LCA 40 MOB - ARDLCA309009C, LCA 40GS MOBILE - ARDLCA309010A, LCA 40GS MOBILE - ARDLCA309010C, LCA 50 MOBILE - ARDLCA309004C, LCA 50 MOBILE B - ARDLCA309005C,

Recall: Z-0759-2024 · Reported January 24, 2024

Enforcement

Recall Number
Z-0759-2024
Event ID
93567
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Getinge Usa Sales Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 24, 2024
Initiation Date
November 8, 2023
Classification Date
January 18, 2024
Address
1 Geoffrey Way, N/A, Wayne, NJ, 07470-2035, United States

Description

Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40 - ARDLCA209021A, LCA 40 - ARDLCA209021C, LCA 40 MOBILE - ARDLCA309009A, LCA 40 WALL - ARDLCA109001A, LCA 40GS - ARDLCA109009A, LCA 40GS - ARDLCA109009C, LCA 100 - ARDLCA109005A, LCA 100 - ARDLCA109005C, LCA 100 - ARDLCA219000A, LCA 100 MOBILE - ARDLCA309006A, LCA 100 MOBILE B - ARDLCA309007A, LCA 100 SAT - ARDSAT209004A, LCA 100 SAT - ARDSAT209005A, LCA 100 V - ARDLCA219000C, LCA 100DF - ARDLCA109008A, LCA 100DF - ARDLCA109008C, LCA 100DF - ARDLCA219003A, LCA 100DF - ARDSAT209008A, LCA 100DF - ARDSAT209009A, LCA 100DF V - ARDLCA219003C, LCA 50 - ARDLCA109004A, LCA 50 - ARDLCA109004C, LCA 50 - ARDLCA209012A, LCA 50 - ARDLCA209012C, LCA 50 - ARDSAT209002A, LCA 50 MOBILE - ARDLCA309004A, LCA 50 MOBILE B - ARDLCA309005A, LCA 50 SAT - ARDSAT209003A, LCA 50 WALL 11 - ARD2LCA00140A, LCA 50DF - ARDLCA109007A, LCA 50DF - ARDLCA109007C, LCA 50DF - ARDLCA209015A, LCA 50DF - ARDLCA209015C, LCA 50DF - ARDSAT209006A, LCA 50DF - ARDSAT209007A, LCA DUO 100 - ARDLCA219001A, LCA DUO 100 V - ARDLCA219001C, LCA DUO 100+50 - ARDLCA219002A, LCA DUO 100+50 V - ARDLCA219002C, LCA DUO 100+50DF - ARDLCA219005A, LCA DUO 100+50DF V - ARDLCA219005C, LCA DUO 100DF - ARDLCA219004A, LCA DUO 100DF V - ARDLCA219004C, LCA DUO 50 - ARDLCA209013A, LCA DUO 50 - ARDLCA209013C, LCA DUO 50DF - ARDLCA209016A, LCA DUO 50DF - ARDLCA209016C, LCA100 MOBILE UPGRADE - ARDLCA506006A, LCA100 UPGRADE - ARDLCA506004A, LCA100 UPGRADE 24V - ARDLCA506005A, LCA100 WALL UPGRADE - ARDLCA506007A, LCA50 UPGRADE - ARDLCA506000A, LCA50 UPGRADE 24V - ARDLCA506001A, LCA50 UPGRADE WALL MOUNTED - ARDLCA506003A, SAT LCA 100 - ARDSAT209004C, SAT LCA 100 - ARDSAT209005C, SAT LCA 100DF - ARDSAT209008C, SAT LCA 100DF - ARDSAT209009C, SAT LCA 50 - ARDSAT209002C, SAT LCA 50 - ARDSAT209003C, SAT LCA 50DF - ARDSAT209006C, SAT LCA 50DF - ARDSAT209007C, LCA 10 MOB - ARDLCA309000A, LCA 10 MOB - ARDLCA309000C, LCA 10 MOB - ARDLCA309002A, LCA 10 MOB - ARDLCA309002C, LCA 10 MOB - ARDLCA309008C, LCA 100 MOBILE - ARDLCA309006C, LCA 100 MOBILE B - ARDLCA309007C, LCA 40 MOB - ARDLCA309001A, LCA 40 MOB - ARDLCA309001C, LCA 40 MOB - ARDLCA309003A, LCA 40 MOB - ARDLCA309003C, LCA 40 MOB - ARDLCA309009C, LCA 40GS MOBILE - ARDLCA309010A, LCA 40GS MOBILE - ARDLCA309010C, LCA 50 MOBILE - ARDLCA309004C, LCA 50 MOBILE B - ARDLCA309005C,

Reason

A potential for a light system to fall in the operating room.

Code Info

All systems affected UDI-DI: 3700712415921 3700712415853 3700712415846 3700712415877 3700712415938 3700712415860 3700712400750 3700712400613 3700712400712 3700712400729 3700712401115 3700712401122 03700712400767 3700712400620 3700712401139 3700712401146 3700712400774 3700712400637 3700712401153 3700712400736 3700712400743 3700712401160 03700712400781 3700712400644 3700712401177 3700712401184 3700712400651 3700712400668 3700712400675 3700712400682 3700712400699 03700712400705 03700712404710 03700712404697 03700712404703 03700712404727 03700712404666 03700712404673 03700712404680

Distribution

US Nationwide. Global Distribution.

Quantity

236793 units in total