PKG, 5MM PEEK HANDLE, 45CM, P/N 0250181164 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Enforcement
- Recall Number
- Z-0757-2015
- Event ID
- 69838
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Endoscopy
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 24, 2014
- Initiation Date
- November 17, 2014
- Classification Date
- December 17, 2014
- Termination Date
- July 17, 2015
- Address
- 5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States
Description
PKG, 5MM PEEK HANDLE, 45CM, P/N 0250181164 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
01292013D 0247065C 0341845C 0342941C 0345124C 0345125C 4082014 0540391C 0540392CD 0542893D 0647218D 07122013E 0743480D 0744062D 0843047D 0845027D 0845859D 0846783D 0940144D 0940148D 0940821D 0941601D 0943110D 0943251D 0943541D 0944721D 0945231D 0945465D 1040630D 1040806D 1041063D 1041468D 1042102D 1043037D 1043039D 1044650D 1045037D 1045055D 1045227D 1045481D 1045638D 1045841D 1046702D 11082010A 1141562 1141562D 1141969D 1141970D 1142168D 1143113D 1143150D 1144236D 1144341D 1144655D 1145341D 1145632D 1145809D 1145939 1145939D 1146669D 1146671D 12122013 1241797D 1241938D 1243268D 1243269D 1244956 1244956D 1245522 1245798 1245799 1245799H 1245800 1245977 1340938 1341257 1341872 1343540 1343540H 1343541 1343542 1344000 1344000H 1344251 1440346 1440346H 1440347 1440347H 1440792 1442304 1442799
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
180,573 units total