ADVIA Chemistry XPT Systems, Software Version 1.0.3, SMN 11127538 Clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour.
Enforcement
- Recall Number
- Z-0755-2016
- Event ID
- 72391
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 10, 2016
- Initiation Date
- October 7, 2015
- Classification Date
- February 4, 2016
- Termination Date
- May 22, 2018
- Address
- 511 Benedict Ave, N/A, Tarrytown, NY, 10591-5005, United States
Description
ADVIA Chemistry XPT Systems, Software Version 1.0.3, SMN 11127538 Clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour.
Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is run, Control Definition screen assumes range defined is 2 SD, Laboratory Information System (LIS) communication/Laboratory Automation (LAS) issue, Printer Driver Resets, ISE Calibration Ranges are too conservative for Urine Sodium, Archiving and deletion may fail and Workstation services may restart.
Software Version 1.0.3
Worldwide Distribution: US (nationwide) in states of: AZ, CA, MA, NY, TX & WA, and to countries of: Bahrain, Egypt, Slovakia, France, Spain, Italy, Norway, Singapore, New Zealand, Germany, Australia, Republic Korea, Netherlands, Denmark, Finland & United Kingdom.
87 units