cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clinical use

Recall: Z-0752-2021 · Reported January 6, 2021

Enforcement

Recall Number
Z-0752-2021
Event ID
86897
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Diagnostics Operations, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
January 6, 2021
Initiation Date
November 23, 2020
Classification Date
December 29, 2020
Termination Date
July 6, 2023
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clinical use

Reason

During the subsequent pipetting (1st or 2nd measurement), the residual reagent can sporadically be transferred from the conus into the reaction cuvette and lead to inaccurately low results for the affected TDMs (Therapeutic Drug Monitoring).

Code Info

Catalog Number 08445605190 08445567190 08253153190 08058580190

Distribution

Domestic: IL, FL, GA, WV

Quantity

152 kits and 7 units