FDA Enforcement
Class II
Terminated
cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clinical use
Recall: Z-0752-2021
·
Reported January 6, 2021
Enforcement
- Recall Number
- Z-0752-2021
- Event ID
- 86897
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- January 6, 2021
- Initiation Date
- November 23, 2020
- Classification Date
- December 29, 2020
- Termination Date
- July 6, 2023
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clinical use
Reason
During the subsequent pipetting (1st or 2nd measurement), the residual reagent can sporadically be transferred from the conus into the reaction cuvette and lead to inaccurately low results for the affected TDMs (Therapeutic Drug Monitoring).
Code Info
Catalog Number 08445605190 08445567190 08253153190 08058580190
Distribution
Domestic: IL, FL, GA, WV
Quantity
152 kits and 7 units