- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615, United States
Description
Giraffe and Panda Bag and Mask Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants.
Reason
Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
Code Info
Model M1139612
Distribution
Worldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA, MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX, VT, WA, WI and WV; and the countries of Australia, Belgium, Canada, China, France, New Zealand, Poland, Saudi Arabia, Switzerland and United Kingdom.
Quantity
276 units total, 109 of those nationwide.