FDA Enforcement Class II Ongoing

20IN (50.8CM) BRAIDED INJ LINE, ROTATING ADAPT, List Number MX682BR

Recall: Z-0745-2024 · Reported January 24, 2024

Enforcement

Recall Number
Z-0745-2024
Event ID
93506
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical Asd Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 24, 2024
Initiation Date
November 10, 2023
Classification Date
January 17, 2024
Address
6250 Shier Rings Rd, N/A, Dublin, OH, 43016-1270, United States

Description

20IN (50.8CM) BRAIDED INJ LINE, ROTATING ADAPT, List Number MX682BR

Reason

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Code Info

Lot Number: 4398428 4404690

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, UAE.

Quantity

31,685 total