ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in vitro diagnostic determination of lactic acid in plasma by colorimetry. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
Enforcement
- Recall Number
- Z-0741-2016
- Event ID
- 73033
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Horiba Instruments Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 10, 2016
- Initiation Date
- January 11, 2016
- Classification Date
- February 2, 2016
- Termination Date
- June 7, 2016
- Address
- 9755 Research Dr, N/A, Irvine, CA, 92618-4626, United States
Description
ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in vitro diagnostic determination of lactic acid in plasma by colorimetry. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
All Lots
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.