FDA Enforcement
Class I
Terminated
HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics
Recall: Z-0741-2013
·
Reported February 6, 2013
Enforcement
- Recall Number
- Z-0741-2013
- Event ID
- 64117
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hamilton Medical Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 6, 2013
- Initiation Date
- December 19, 2012
- Classification Date
- January 31, 2013
- Termination Date
- April 22, 2014
- Address
- 4990 Energy Way, P O Box 30008, Reno, NV, 89502-4123, United States
Description
HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics
Reason
Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.
Code Info
Software Versions 1.1.2 and below; Part numbers 161005, 161006.
Distribution
US Nationwide Distribution including the country of Switzerland.
Quantity
97 total ventilators; 69 of part number 161005 and 28 of part number 161006