FDA Enforcement Class II Terminated

7% BSA (Bovine Serum Albumin), Product Code: 8262487 - Product Usage: is used to dilute samples when assay values exceed the reportable (dynamic) range using the VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Recall: Z-0739-2020 · Reported January 1, 2020

Enforcement

Recall Number
Z-0739-2020
Event ID
84123
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho Clinical Diagnostics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
January 1, 2020
Initiation Date
October 14, 2019
Classification Date
December 26, 2019
Termination Date
November 13, 2020
Address
1001 US Highway 202, Raritan, NJ, 08869-1424, United States

Description

7% BSA (Bovine Serum Albumin), Product Code: 8262487 - Product Usage: is used to dilute samples when assay values exceed the reportable (dynamic) range using the VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Reason

Potential for the instability of the diluents VITROS 7% BSA (Product Code 8262487) and VITROS Urine Electrolyte Diluent (UED) (Product Code 1110352) when stored in a 2mL cup for 7 days

Code Info

Lot with Expiry Date: D6289 2019-11-20 E6382 2020-01-15 F6602 2020-05-02 G6679 2020-08-13 H6897 2020-10-01 J7129 2021-01-23 K7421 2021-07-01 L7592 2021-08-26

Distribution

Worldwide distribution - US Nationwide distribution and countries BM, AU, BR, CA, CL, CN, CO, IN, JP, MX, SG, GB, FR, DE, IT, ES, PT, PL, RU, DK, NO, SE, BE, NL.

Quantity

16,584