ABX PENTRA Enzymatic Creatinine CP ref. A11A01907 reagent is intended for the quantitative in vitro diagnostic determination of creatinine in human serum, plasma and urine based on an enzymatic method of using a multi-step approach ending with a photometric end-point reaction.
Enforcement
- Recall Number
- Z-0739-2016
- Event ID
- 73033
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Horiba Instruments Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 10, 2016
- Initiation Date
- January 11, 2016
- Classification Date
- February 2, 2016
- Termination Date
- June 7, 2016
- Address
- 9755 Research Dr, N/A, Irvine, CA, 92618-4626, United States
Description
ABX PENTRA Enzymatic Creatinine CP ref. A11A01907 reagent is intended for the quantitative in vitro diagnostic determination of creatinine in human serum, plasma and urine based on an enzymatic method of using a multi-step approach ending with a photometric end-point reaction.
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
All Lots
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.