Ongoing
NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 assay strips in individual pouches. Box, pouch and strips all labeled with product name, ref #, lot #, and expiration date.
Recall: Z-0738-2022
·
Reported March 16, 2022
Enforcement
- Recall Number
- Z-0738-2022
- Event ID
- 89675
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- NeuMoDx Molecular Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- March 16, 2022
- Initiation Date
- January 21, 2022
- Classification Date
- March 8, 2022
- Address
- 1250 Eisenhower Pl, N/A, Ann Arbor, MI, 48108-3281, United States
Description
NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 assay strips in individual pouches. Box, pouch and strips all labeled with product name, ref #, lot #, and expiration date.
Reason
A risk for product contamination was identified for the affected product lot. If used, there is an increased potential for SARS-CoV-2 false positive results when using the product.
Code Info
Lot: 116305 S/L-02 (Sublot 02),UDI: 10814278020830
Distribution
US Nationwide distribution in the states of FL, MD, MI, NY.
Quantity
102 boxes