Custom procedure packs, containing banded 4x4 Raytec 16ply gauze sponges. 1) Regard, PACEMAKER, Item 800221010, Lot # 82466. 2) Regard, BASIC, Item 800109003 Lot # 83411. 3) Regard, MAJOR EAR, Item 800070009, Lot # 82442. 4) Regard, NEURO, Item 800040011, Lot # 82504. 5) Regard, LAP CHOLE, Item 800003011, Lot # 82497. 6) Regard, SHOULDER ARTHRO, Item 800075010, Lot # 82490. 7) Regard, CERVICAL, Item 800058007, Lot # 82592. 8) Regard, ARTHROSCOPY, Item 800027008, Lot # 82582. 9) Regard, MAJOR SURGERY, Item 800013010, Lot # 82844. 10) Regard, GENERAL SURGERY, Item 800057012, Lot # 82832. 11) Regard, SEPTO SINUS ENDO, Item 800024006, Lot # 82876. 12) Regard, MINOR FMOL-LADY OF THE LAKE, Item 800019009, Lot # 82893. 13) Regard, KNEE ARTHRO, Item 800072009, Lot # 83058. 14) Regard, LAP, Item 800096005, Lot # 83139. 15) Regard, KNEE ARTHROSCOPY, Item 800121, Lot # 83251. 16) Regard, CYSTOSCOPY, Item 800093007, Lot # 83138. 17) Regard, LAP CHOLE, Item 800035007, Lot # 83257. 18) Regard, PERIPHERAL VASCULAR, Item 800007011, Lot # 83106. 19) Regard, PACEMAKER, Item 800083006, Lot # 82886. 20) Regard, I&D, Item 800056005, Lot # 83135. 21) Regard, HEAD & NECK, Item 800068010, Lot # 83050. 22) Regard, ROBOTICS, Item 800103007, Lot # 83264. 23) Regard, MAJOR EAR, Item 800070010, Lot # 82879. 24) Regard, ORTHO EXTREMITY, Item 800042009, Lot # 83383. 25) Regard, CRANIOTOMY, Item 800025015, Lot # 83271. 26) Regard, MINOR ORTHO, Item 800020007, Lot # 83583. 27) Regard, VATS, Item 800118001, Lot # 83421. 28) Regard, ARTHROSCOPY, Item 800027008, Lot # 83560. 29) Regard, OPEN HEART, Item 800023010, Lot # 83273. 30) Regard, PERIPHERAL VASCULAR, Item 800018011, Lot # 83575. 31) Regard, GENERAL PROCEDURE, Item 800000008, Lot # 83576. 32) Regard, DR WHITE OPEN HRT A&B, Item 800045012, Lot # 82847. 33) Regard, NEURO, Item 800008008, Lot # 83598. 34) Regard, LAPAROSCOPY, Item 880320008, Lot # 82472. 35) Regard, PACEMAKER, Item 880074023, Lot # 82417. 36) Regard, THYROID, Item 880425002, Lot # 82389. 37) Regard, LAPAROSCOPY, Item 880350012, Lot # 82473. 38) Regard, CHEST, Item 880373005, Lot # 82427. 39) Regard, BASIC CATARACT, Item 880348006, Lot # 82501. 40) Regard, CCOR HAND, Item 880337009, Lot # 82496. 41) Regard, CRANIOTOMY, Item 880396004, Lot # 82507. 42) Regard, TUR CYSTO, Item 880357006, Lot # 82574. 43) Regard, TOTAL HIP, Item 880266, Lot # 82478. 44) Regard, DAVINCI, Item 880132021, Lot # 82469. 45) Regard, KNEE ARTHROSCOPY, Item 880368, Lot # 82485. 46) Regard, UROLOGY, Item 880306010, Lot # 82584. 47) Regard, MINOR BREAST, Item 880195017, Lot # 82588. 48) Regard, LABOR & DELIVERY, Item 880081018, Lot # 82600. 49) Regard, PELVIC PROCEDURE, Item 880214016, Lot # 82597. 50) Regard, CCOR MINOR, Item 880336010, Lot # 82593. 51) Regard, YOUNG LUMBAR, Item 880432005, Lot # 82594. 52) Regard, PACEMAKER, Item 880418003, Lot # 82606. 53) Regard, LOCAL, Item 880087014, Lot # 82828. 54) Regard, COLON LITHOTOMY, Item 880395002, Lot # 82825. 55) Regard, EYE, Item 880099029, Lot # 82882. 56) Regard, FESS, Item 880404001, Lot # 83033. 57) Regard, CYSTO, Item 880312009, Lot # 82880. 58) Regard, EXTREMITY PACK, Item 880167018, Lot # 82887. 59) Regard, PHERIPHERAL VASC, Item 880077, Lot # 83048. 60) Regard, T & A, Item 880423002, Lot # 83073. 61) Regard, C CABG A & B, Item 880080, Lot # 83042. 62) Regard, DR MARBARGER, Item 880079011, Lot # 83124. 63) Regard, MPR LOCAL, Item 880360006, Lot # 83133. 64) Regard, KNEE ARTHROSCOPY, Item 880268, Lot # 83095. 65) Regard, MINOR PROCEDURE, Item 880261009, Lot # 83132. 66) Regard, EXTREMITY, Item 880262009, Lot # 83071. 67) Regard, LAPAROSCOPY, Item 880102018, Lot # 83238. 68) Regard, GENERAL LAPAROSCOPY, Item 880405002, Lot # 83254. 69) Regard, MAJOR PROCEDURE, Item 880270008, Lot # 83240. 70) Regard, YOUNG LUMBAR, Item 880432005, Lot # 83232. 71) Regard, MINOR LITHOTOMY, Item 88

Recall: Z-0737-2021 · Reported January 6, 2021

Enforcement

Recall Number
Z-0737-2021
Event ID
86830
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ROi CPS LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 6, 2021
Initiation Date
November 9, 2020
Classification Date
December 28, 2020
Termination Date
May 13, 2021
Address
3000 E Sawyer Rd, N/A, Republic, MO, 65738-2758, United States

Description

Custom procedure packs, containing banded 4x4 Raytec 16ply gauze sponges. 1) Regard, PACEMAKER, Item 800221010, Lot # 82466. 2) Regard, BASIC, Item 800109003 Lot # 83411. 3) Regard, MAJOR EAR, Item 800070009, Lot # 82442. 4) Regard, NEURO, Item 800040011, Lot # 82504. 5) Regard, LAP CHOLE, Item 800003011, Lot # 82497. 6) Regard, SHOULDER ARTHRO, Item 800075010, Lot # 82490. 7) Regard, CERVICAL, Item 800058007, Lot # 82592. 8) Regard, ARTHROSCOPY, Item 800027008, Lot # 82582. 9) Regard, MAJOR SURGERY, Item 800013010, Lot # 82844. 10) Regard, GENERAL SURGERY, Item 800057012, Lot # 82832. 11) Regard, SEPTO SINUS ENDO, Item 800024006, Lot # 82876. 12) Regard, MINOR FMOL-LADY OF THE LAKE, Item 800019009, Lot # 82893. 13) Regard, KNEE ARTHRO, Item 800072009, Lot # 83058. 14) Regard, LAP, Item 800096005, Lot # 83139. 15) Regard, KNEE ARTHROSCOPY, Item 800121, Lot # 83251. 16) Regard, CYSTOSCOPY, Item 800093007, Lot # 83138. 17) Regard, LAP CHOLE, Item 800035007, Lot # 83257. 18) Regard, PERIPHERAL VASCULAR, Item 800007011, Lot # 83106. 19) Regard, PACEMAKER, Item 800083006, Lot # 82886. 20) Regard, I&D, Item 800056005, Lot # 83135. 21) Regard, HEAD & NECK, Item 800068010, Lot # 83050. 22) Regard, ROBOTICS, Item 800103007, Lot # 83264. 23) Regard, MAJOR EAR, Item 800070010, Lot # 82879. 24) Regard, ORTHO EXTREMITY, Item 800042009, Lot # 83383. 25) Regard, CRANIOTOMY, Item 800025015, Lot # 83271. 26) Regard, MINOR ORTHO, Item 800020007, Lot # 83583. 27) Regard, VATS, Item 800118001, Lot # 83421. 28) Regard, ARTHROSCOPY, Item 800027008, Lot # 83560. 29) Regard, OPEN HEART, Item 800023010, Lot # 83273. 30) Regard, PERIPHERAL VASCULAR, Item 800018011, Lot # 83575. 31) Regard, GENERAL PROCEDURE, Item 800000008, Lot # 83576. 32) Regard, DR WHITE OPEN HRT A&B, Item 800045012, Lot # 82847. 33) Regard, NEURO, Item 800008008, Lot # 83598. 34) Regard, LAPAROSCOPY, Item 880320008, Lot # 82472. 35) Regard, PACEMAKER, Item 880074023, Lot # 82417. 36) Regard, THYROID, Item 880425002, Lot # 82389. 37) Regard, LAPAROSCOPY, Item 880350012, Lot # 82473. 38) Regard, CHEST, Item 880373005, Lot # 82427. 39) Regard, BASIC CATARACT, Item 880348006, Lot # 82501. 40) Regard, CCOR HAND, Item 880337009, Lot # 82496. 41) Regard, CRANIOTOMY, Item 880396004, Lot # 82507. 42) Regard, TUR CYSTO, Item 880357006, Lot # 82574. 43) Regard, TOTAL HIP, Item 880266, Lot # 82478. 44) Regard, DAVINCI, Item 880132021, Lot # 82469. 45) Regard, KNEE ARTHROSCOPY, Item 880368, Lot # 82485. 46) Regard, UROLOGY, Item 880306010, Lot # 82584. 47) Regard, MINOR BREAST, Item 880195017, Lot # 82588. 48) Regard, LABOR & DELIVERY, Item 880081018, Lot # 82600. 49) Regard, PELVIC PROCEDURE, Item 880214016, Lot # 82597. 50) Regard, CCOR MINOR, Item 880336010, Lot # 82593. 51) Regard, YOUNG LUMBAR, Item 880432005, Lot # 82594. 52) Regard, PACEMAKER, Item 880418003, Lot # 82606. 53) Regard, LOCAL, Item 880087014, Lot # 82828. 54) Regard, COLON LITHOTOMY, Item 880395002, Lot # 82825. 55) Regard, EYE, Item 880099029, Lot # 82882. 56) Regard, FESS, Item 880404001, Lot # 83033. 57) Regard, CYSTO, Item 880312009, Lot # 82880. 58) Regard, EXTREMITY PACK, Item 880167018, Lot # 82887. 59) Regard, PHERIPHERAL VASC, Item 880077, Lot # 83048. 60) Regard, T & A, Item 880423002, Lot # 83073. 61) Regard, C CABG A & B, Item 880080, Lot # 83042. 62) Regard, DR MARBARGER, Item 880079011, Lot # 83124. 63) Regard, MPR LOCAL, Item 880360006, Lot # 83133. 64) Regard, KNEE ARTHROSCOPY, Item 880268, Lot # 83095. 65) Regard, MINOR PROCEDURE, Item 880261009, Lot # 83132. 66) Regard, EXTREMITY, Item 880262009, Lot # 83071. 67) Regard, LAPAROSCOPY, Item 880102018, Lot # 83238. 68) Regard, GENERAL LAPAROSCOPY, Item 880405002, Lot # 83254. 69) Regard, MAJOR PROCEDURE, Item 880270008, Lot # 83240. 70) Regard, YOUNG LUMBAR, Item 880432005, Lot # 83232. 71) Regard, MINOR LITHOTOMY, Item 88

Reason

Gauze sponges contained in custom kits may be flaking or fraying.

Code Info

Lot Numbers: 82466, 83411, 82442, 82504, 82497, 82490, 82592, 82582, 82844, 82832, 82876, 82893, 83058, 83139, 83251, 83138, 83257, 83106, 82886, 83135, 83050, 83264, 82879, 83383, 83271, 83583, 83421, 83560, 83273, 83575, 83576, 82847, 83598, 82472, 82417, 82389, 82473, 82427, 82501, 82496, 82507, 82574, 82478, 82469, 82485, 82584, 82588, 82600, 82597, 82593, 82594, 82606, 82828, 82825, 82882, 83033, 82880, 82887, 83048, 83073, 83042, 83124, 83133, 83095, 83132, 83071, 83238, 83254, 83240, 83232, 83244, 82872, 83222, 82831, 83263, 83111, 83267, 83261, 83274, 83258, 83266, 83388, 83424, 83417, 83426, 83415, 83427, 83557, 83536, 83540, 83418, 83581, 83550, 83428, 83579, 83296, 83268, 83752, 82591, 82589, 82640, 82813, 82513, 82596, 83069, 82852, 83252, 83101, 83293, 83392, 83572, 83120, 83584, 83422, 83955, 82399, 82474, 82403, 82492, 82509, 82470, 82465, 82491, 82575, 82576, 82521, 82525, 82481, 82577, 82500, 82601, 82607, 82595, 82605, 82519, 82523, 82603, 82625, 82888, 83045, 82855, 83066, 82849, 83036, 83049, 82890, 83242, 83235, 83115, 83064, 83046, 83118, 83245, 83126, 83377, 83134, 83389, 83127, 83286, 83269, 83398, 83292, 83285, 83366, 83423, 83391, 83136, 83445, 83283, 83561, 83545, 83585, 83589, 83580, 83944, 83771.

Distribution

US Domestic Distribution Only: MO, NC, LA and FL.

Quantity

17,700 total kits