FDA Enforcement Class II Terminated

HemosIL Plasminogen, Part No. 0020009000, IVD. Instrumentation Laboratory Company. Kit contents: 2 x 2.5 mL Streptokinase reagent and 2 x 2 mL Chromogenic substrate

Recall: Z-0736-2016 · Reported February 10, 2016

Enforcement

Recall Number
Z-0736-2016
Event ID
72972
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Instrumentation Laboratory Co.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
February 10, 2016
Initiation Date
December 18, 2015
Classification Date
February 1, 2016
Termination Date
August 30, 2017
Address
180 Hartwell Rd, N/A, Bedford, MA, 01730-2443, United States

Description

HemosIL Plasminogen, Part No. 0020009000, IVD. Instrumentation Laboratory Company. Kit contents: 2 x 2.5 mL Streptokinase reagent and 2 x 2 mL Chromogenic substrate

Reason

This Lot is not meeting labeled on-board instrument stability claims of 5 days.

Code Info

Lot No. N0455811, Expiration Date 03/31/2017

Distribution

Worldwide Distribution - US Distribution and to the countries of : Canada, Argentina, Austria, Belgium, Brazil, China, Czech Republic, France, Germany, Hong Kong, Italy, Lithuania, Panama, Poland, Russia, Saudi Arabia, United Kingdom, and Venezuela.

Quantity

Domestic: 108 kits, Foreign: 197 kits