FDA Enforcement Class II Completed

FreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 Operating System

Recall: Z-0734-2024 · Reported January 24, 2024

Enforcement

Recall Number
Z-0734-2024
Event ID
93393
Classification
Class II
Status
Completed
Product Type
Devices
Firm
Abbott Diabetes Care, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 24, 2024
Initiation Date
February 9, 2023
Classification Date
January 16, 2024
Address
1360 S Loop Rd, N/A, Alameda, CA, 94502-7000, United States

Description

FreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 Operating System

Reason

If using affected glucose monitoring app on Android 13 Operating System, extended periods of signal loss may be experienced, due to app not connected, which could impact ability to receive glucose reading/alarms, which could lead to undetected low or high glucose, which could result in delayed treatment: not taking insulin (for high glucose), or not taking glucose (for low glucose) when required.

Code Info

UDI-DI: 00357599816001, Software v.3.4.0 and below

Distribution

US Nationwide distribution in the states of NY, CA, NJ, UT, PA, MS, OH, TX, MN, FL, LA, NE, MI, VA, NC, MA, IN, SC, IA.

Quantity

45,173