FDA Enforcement Class II Terminated

PREMIER Cryptococcal Antigen, Catalog Number 602096 Product Usage: The PREMIER Cryptococcal Antigen enzyme immunoassay (EIA) is a screening or a semi-quantitative test system for the detection of capsular polysaccharide antigens of Cryptococcus neoformans in serum and cerebrosoinal fluid (CSF).

Recall: Z-0732-2019 · Reported January 23, 2019

Enforcement

Recall Number
Z-0732-2019
Event ID
81787
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Meridian Bioscience Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 23, 2019
Initiation Date
December 17, 2018
Classification Date
January 14, 2019
Termination Date
April 30, 2020
Address
3471 River Hills Dr, Cincinnati, OH, 45244-3023, United States

Description

PREMIER Cryptococcal Antigen, Catalog Number 602096 Product Usage: The PREMIER Cryptococcal Antigen enzyme immunoassay (EIA) is a screening or a semi-quantitative test system for the detection of capsular polysaccharide antigens of Cryptococcus neoformans in serum and cerebrosoinal fluid (CSF).

Reason

An enzyme reagent included in the kit is not maintaining stability through claimed product expiration, resulting in the potential for false negative results when tested with patient specimens and the positive control.

Code Info

Lot Numbers: 602096K089, 602096K090, 602096K091

Distribution

Worldwide Distribution - US Nationwide in the states of: AK, CA, CO, DE, IL, KS, KY, LA, MD, MO, NC, ND, NJ, NY, SC, TX, and VA. The products were distributed to the following foreign countries: India and Italy.

Quantity

184