FDA Enforcement
Class I
Terminated
ultraView SISH Detection Kit, Ventana Part Number 780-001, Roche GMMI 05271967001 in vitro diagnostic
Recall: Z-0732-2018
·
Reported March 7, 2018
Enforcement
- Recall Number
- Z-0732-2018
- Event ID
- 78942
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ventana Medical Systems Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 7, 2018
- Initiation Date
- December 19, 2017
- Classification Date
- September 14, 2018
- Termination Date
- November 22, 2021
- Address
- 1910 E Innovation Park Dr, N/A, Oro Valley, AZ, 85755-1962, United States
Description
ultraView SISH Detection Kit, Ventana Part Number 780-001, Roche GMMI 05271967001 in vitro diagnostic
Reason
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
Code Info
UDI 04015630972296, Lot Number: Y15133
Distribution
Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
Quantity
6 units