FDA Enforcement
Class II
Ongoing
VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
Recall: Z-0728-2023
·
Reported December 28, 2022
Enforcement
- Recall Number
- Z-0728-2023
- Event ID
- 91174
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Icotec Ag
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 28, 2022
- Initiation Date
- November 3, 2022
- Classification Date
- December 19, 2022
- Address
- Industriestrasse 12, Altstatten Sg, N/A, Switzerland
Description
VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
Reason
During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Code Info
UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045
Distribution
US Nationwide distribution in the state of GA.
Quantity
6 devices