FDA Enforcement
Class II
Ongoing
1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope/endoscope/arthroscope/sinuscope is used.
Recall: Z-0728-2022
·
Reported March 16, 2022
Enforcement
- Recall Number
- Z-0728-2022
- Event ID
- 89515
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 16, 2022
- Initiation Date
- January 14, 2022
- Classification Date
- March 7, 2022
- Address
- 5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States
Description
1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope/endoscope/arthroscope/sinuscope is used.
Reason
A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.
Code Info
All serial numbers of 1688 CCU, UDI: 07613327420081, with Software Revisions: 3.0.6, 4.0.13, and 4.0.18
Distribution
Worldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL, IN, FR, ES, HK, IT, GB, CO, PE, MY, TH, CA, JP, MX, AR, NP, SG, PH, ZA, CL, KR, CN
Quantity
11,955