S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The SMART Control stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The stent is equipped with 12 tantalum markers, 6 at each end, to increase the visibility of the stent under fluoroscopic imaging. The SMART Control stent system is designed to be delivered through an appropriately sized catheter sheath introducer.
Enforcement
- Recall Number
- Z-0728-2020
- Event ID
- 84228
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Cardinal Health Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 1, 2020
- Initiation Date
- November 14, 2019
- Classification Date
- December 20, 2019
- Address
- 7000 Cardinal Pl, Dublin, OH, 43017-1091, United States
Description
S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The SMART Control stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The stent is equipped with 12 tantalum markers, 6 at each end, to increase the visibility of the stent under fluoroscopic imaging. The SMART Control stent system is designed to be delivered through an appropriately sized catheter sheath introducer.
Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
All Lot Codes. Affected SKUs: N1020ABR, N1020BBR, N1020TBR, N1040ABR, N1040BBR, N1040EBR, N1040TBR, N1060ABR, N1060BBR, N1060EBR, N1060TBR, N1080ABR, N1080BBR, N1080EBR, N1080TBR, N6120BV, N620ABR, N620BBR, N620TBR, N640ABR, N640BBR, N640TBR, N660ABR, N660BBR, N660TBR, N680ABR, N680BBR, N680TBR, N7120BV, N720ABR, N720BBR, N720TBR, N740ABR, N740BBR, N740TBR, N760ABR, N760BBR, N760TBR, N780ABR, N780BBR, N780TBR, N8120BV, N820ABR, N820BBR, N820TBR, N840ABR, N840BBR, N840EBR, N840TBR, N860ABR, N860BBR, N860EBR, N860TBR, N880ABR, N880BBR, N880EBR, N880TBR, N920ABR, N920BBR, N920TBR, N940ABR, N940BBR, N940TBR, N960ABR, N960BBR, N960TBR, N980ABR, N980BBR, N980TBR.
Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saint-Barthelemy, Saudia Arabia, Sinapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela.
267,247 units total (262,641 OUS, 4,606 US)