FDA Enforcement Class II Terminated

PKG, 5MM INSERT, ENDO GRASPER, 33CM, P/N 0250080699 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Recall: Z-0727-2015 · Reported December 24, 2014

Enforcement

Recall Number
Z-0727-2015
Event ID
69838
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 24, 2014
Initiation Date
November 17, 2014
Classification Date
December 17, 2014
Termination Date
July 17, 2015
Address
5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States

Description

PKG, 5MM INSERT, ENDO GRASPER, 33CM, P/N 0250080699 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code Info

844411 844628 844811 846760 940456 943679 946279 1042100 1043033 1043101 1045370 1140297 1141155 1142927 1144484 1145067 1145725 1146838 1242927 1244166 1245602 1340116 1341541 1342757 1342758 1344780 1346663 544589

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Quantity

180,573 units total