FDA Enforcement Class II Ongoing

PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870

Recall: Z-0726-2024 · Reported January 24, 2024

Enforcement

Recall Number
Z-0726-2024
Event ID
93668
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Limacorporate S.p.A
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
January 24, 2024
Initiation Date
August 29, 2023
Classification Date
January 12, 2024
Address
Via Nazionale 52, Di San Daniele, San Daniele Del Friuli, N/A, N/A, Italy

Description

PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870

Reason

Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.

Code Info

Product Code/UDI-DI: 1975.14.500/08033390244824, 1975.14.800/08033390244893, 1975.14.510/08033390231299, 1975.14.810/08033390244909, 1975.14.615/08033390244848, 1975.14.815/08033390244916, 1975.14.865/08033390244930, 1975.14.620/08033390231312, 1975.14.820/08033390244923, 1975.14.870/08033390244947

Distribution

US: IL, MO, CA, IN, MI, OK

Quantity

179