FDA Enforcement Class II Terminated

PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 33CM, P/N 0250080694 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Recall: Z-0724-2015 · Reported December 24, 2014

Enforcement

Recall Number
Z-0724-2015
Event ID
69838
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 24, 2014
Initiation Date
November 17, 2014
Classification Date
December 17, 2014
Termination Date
July 17, 2015
Address
5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States

Description

PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 33CM, P/N 0250080694 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code Info

44493 647221 647992 745498 842129 845882 846787 847844 940092 940153 941265 941741 944824 1040124 1040628 1041467 1042526 1042527 1043031 1043779 10441058 1044108 1045039 1045371 1046186 1046395 1046955 1047068 1140260 1141156 1141790 1142614 1144485 1145002 1145148 1145969 1146672 1240949 1241420 1241798 1242459 1243202 1244181 1244438 1244527 1244594 1245525 1245545 1246645 1340475 1341523 1342388 1343130 1343554 1344027 1344278 1344686 1346102 1440693 1441635 1443223

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Quantity

180,573 units total