FDA Enforcement Class II Ongoing

IntelePACS - InteleConnect / TechPortal

Recall: Z-0723-2026 · Reported December 3, 2025

Enforcement

Recall Number
Z-0723-2026
Event ID
97868
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
INTELERAD MEDICAL SYSTEMS INCORPORATED
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 3, 2025
Initiation Date
October 17, 2025
Classification Date
November 26, 2025
Address
800, Boul. de Maisonneuve Est, 12th floor, Montreal, N/A, N/A, Canada

Description

IntelePACS - InteleConnect / TechPortal

Reason

Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.

Code Info

UDI-DI: B228INTELEPACS0. Version/Dates of Distribution: PACS-5-6-1-R64 through R68/02 Jul 2025 - 10 Sep 2025, PACS-5-7-1-R42 through R47/26 June 2025 - 25 Sep 2025, PACS 5-8-1-R20 through R23/26 Jun 2025 - 25 Sep 2025, PACS-5-9-1-R1 through R3/30 Jun 2025 - 09 Sep 2025

Distribution

US Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.

Quantity

19